39
60126903-A
Alarm
Error
Code
Solution
Symbol
Indicator
Light Color
Sound
Indication
High Pressure
H3020
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
High Presure Diagnostics
H302x
Contact provider/Refer to Service
Manual
White
with red
dot
Constant
Audible
Alarm
Low Pressure Diagnostics
H304x
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
High Temperature Shut
Down
H4000
1. Inspect for blocked vents
2. Ensure device is in specified
operating environment
3. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Valve Coil Failure
H5000
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Fan Not Connected
H6000
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Fan Speed Out of Range
H6100
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Invacare® Platinum® 5NXG Oxygen Concentrator