6
Right Park Brake Fault
Ensure brake lever is in the drive position
before turning on the wheelchair. Ensure
motor cable is plugged into the controller.
Contact Invacare/Dealer for service.
7
Remote Fault
Check to make sure joystick is connected
properly. Turn Joystick off then on.
Contact Invacare/Dealer for service.
8
Controller Fault
Contact Invacare/Dealer for service.
9
Communications Fault
Check joystick cable connections. Check
joystick cable and connectors for damage.
Contact Invacare/Dealer for service.
10
General Fault
Contact Invacare/Dealer for service.
11
Incompatible or incorrect Remote
Wrong type of remote connected. Contact
Invacare/Dealer for service.
8.1.5
CMPJ™+, PSR+, PSF+ Joysticks or Displays
SYMPTOM
PROBABLE CAUSE
solutions
LEFT BRAKE FAULT or RIGHT BRAKE
FAULT displays and wheelchair does not drive.
Motor lock levers disengaged (Error code E9
or E10).
Engage motor lock levers. Refer to owner’s
manual shiped with wheelchair.
CHARGER PLUGGED IN displays.
Battery charger connected (Error code E28).
Unplug battery charger from the wheelchair.
Refer to owner’s manual shipped with
wheelchair.
BATTERY FAULT displays and the wheelchair
does not drive.
Batteries need to be charged (Error code E14).
Charge batteries. Refer to owner’s manual
shipped with wheelchair. If batteries fail to
charge properly, check battery charger or
replace batteries.
1143155-R-01
89