Invacare® Formula™ CG Seating
8 Troubleshooting
8.1
Electrical
For additional troubleshooting information and explanation of error codes, refer to the wheelchair base owner’s manual and the individual
Electronics Manual supplied with each wheelchair.
8.1.1
All Joysticks
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Wheelchair Power On but does not drive.
System tilted or reclined beyond drive lock-out
angle (20°).
Motor lock levers are disengaged.
Return to neutral position (upright and
completely lowered).
Contact Invacare/Dealer for service if this
does not solve the problem.
Engage motor lock levers. See wheelchair base
owner’s manual.
Seating system not functioning or working
intermittently.
Low batteries.
Faulty electrical connection.
Seat has been driven under a heavy load for an
extended period of time.
Charge batteries.
Check all connections.
Allow time for the electronics to cool down
(Light Duty Use).
8.1.2
SPJ+, SPJ+ w/PSS or SPJ+ w/ACC Joysticks
The joystick information gauge and the service indicator give indications of the type of fault or error detected by the control module. When
a fault is detected, the wheelchair may stop and not drive. The LEDs on the information gauge may flash in a particular pattern or the service
indicator light will flash. The number or type of flashes indicates the nature of the error. If multiple errors are found, only the first error
encountered by the control module will be displayed.