When installing multi- directional valve block, make 
sure that all bolts are tightened uniformly and 
opposite bolts are tightened in sequence. 

 

7.2 

p-Transmitter: 

For liquid measurements, the differential pressure 
transmitter should always be installed below the 
ITABAR sensor in order to avoid the occurrence of 
air bubbles in the instrument connections (see 
Fig. 13).  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
Fig. 13 

 

When measuring dry gases and air, the differential 
pressure transmitter should always be installed 
above the ITABAR sensor in order to avoid 
measurement degradation due to condensation and 
presence of solid particles (see Fig. 14).

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

Fig. 14 

7.3  ITABAR sensor with flange plate 

By using of the flange plate version, you can 
mounting the 

p-transmitters direct onto the flow 

senor (s. fig. 15). By this version you don`t must 
install differential pressure lines and screwed 
connections. In combination with a multi- 
directional valve block is this version the best 
solution. 
When installing multi- directional valve blocks, 
make sure that all bolts are tightened uniformly and 
opposite bolts are tightened in sequence. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Abb. 14 

Fig. 15 

 

 

 

 

Summary of Contents for Itabar IBF

Page 1: ...Installation of the ITABAR Sensor 6 7 Installation of the Attachments and the Dp transmitter 8 8 Measurement Start Up 9 9 Preventive Maintenance 9 10 Trouble Shooting 9 INTRA AUTOMATION GmbH Me und R...

Page 2: ...ns ITABAR Flow sensors type IBF can be used under the following conditions Operating pressure max 320 bar Operating temperature 50 C to 1000 C Nominal Pipe Diameters DN 40 to DN 12000 The sensor mater...

Page 3: ...low measurements of liquids the ITABAR sensor must be installed in the lower half of the pipe circumference the connections to the instruments have to be located below the pipe axis This way gas or ai...

Page 4: ...hysical principles Its design incorporates no moving parts which are subject to wear Fig 6 The sensor is not affected by being slightly out of alignment The influence on the accuracy of the measuremen...

Page 5: ...ion of the most advantageous installation location As a general rule regulating valves throttle valves and gate type valves should be installed behind the sensor NOTE If the recommended straight pipe...

Page 6: ...flow direction Tighten the bolts and nuts Note On flanges with eight bolt holes the welding stud must be welded on so that the bolt holes in the flange form an angle of 22 5 with the pipe axis see Fi...

Page 7: ...m diameter into the pipe The subsequent Stepps correspond to the installation of the ITABAR sensor type IBF 21 Oberserve the H dimension during the welding see also Fig 9 6 7 Installation of Type IBF...

Page 8: ...pressure transmitter should always be installed above the ITABAR sensor in order to avoid measurement degradation due to condensation and presence of solid particles see Fig 14 Fig 14 7 3 ITABAR senso...

Page 9: ...ng errors occur they may possibly be corrected quite easily error No differential pressure indication Correction Check whether all instrument valves to the p transmitter are opened Check the alignment...

Reviews: