S Y S T E M O P E R A T I O N
CONFIGURE SYSTEM Screen Detailed
ATS-515, -525 & ATS-615 & -625 Interface & Applications Manual
3-31
Temperature
Performance
Auto: temperature performance is set to auto
Heat Only: The compressor is shut off. The system only drives to
hot temperatures.
Off: compressor is toggled on, the system heats and cools.
DEFROST
Set to On to display
Defrost Screen
countdown timer and to enter
Defrost mode.
Head always up
If On:
-at test start: Head remains up, air flow is on
-at test end: Head remains up, air flow is off
If Off:
-at test start: Head goes down, air flow is on
-at test end: Head goes up, air flow is off
Line Voltage Selection
Displays the voltage and frequency required by the system.
Part