S Y S T E M O P E R A T I O N
PROGRAM TEST SETUP Screen
3-20
ATS-515, -525 & ATS-615 & -625 Interface & Applications Manual
SELECT
program button
Press to cycle through the TEST SETUPS “0” - “12”.
ATTENTION
- Parameter changes and Test Setups should be saved
under Test Setup “1” through “12”.
- DO NOT save under Test Setup “0” as “0” is the
System default. Test Setup “0” will be over written
because it always reflects the last Test Setup used.
- Test Setup “0” through “12” can be selected with
the
“SELECT”
program button while in the Setup
screens.
DEFINE
program button
Selects the
DEFINE PARAMETERS
Screen
. This screen allows
the user to define test setup parameters that are common to both
Manual and Program Test Setups.
For greater detail, see
28.
DONE
program button
Press to
SAVE
configured settings and return to the
PROGRAM
TEST SETUP
Screen.
RAMP °C/minute
The rate (in °C/minute) at which the
THERMOSTREAM
®
ramps
up/down to the corresponding Setpoint (SetP°C).
This parameter may be changed using the Encoder.
• “9999” is the maximum value that may be entered. The
THERMOSTREAM
®
ramps up/down at its fastest rate.
• “0” is the minimum value that may be entered. The
THERMOSTREAM
®
will NOT cycle up/down to the
corresponding Setpoint (SetP°C). A setting of “0” effectively
disables the Setpoint for cycling purposes.
Row #
The profile for each unique setpoint is filed in rows numbered “0”
through “12”.
SetP°C
Setpoint (temperature °C) for each respective row.
This parameter may be changed using the Encoder.
Wndw °C
Set range above/below setpoint, in which system is at setpoint.
This parameter may be changed using the Encoder.
Example:
a 50 °C setpoint, with 1.0 °C Window, is at setpoint
between 49 °C to 51 °C.
TEST SETUPS 0 - 12
Use the
“SELECT”
program button to cycle through the TEST
SETUPS.
Soak Sec.
SOAK TIME in seconds for the corresponding Setpoint (SetP°C).
• Soak time is the time system remains at setpoint, within Wndw
°C, before the "At Temp" test time begins.
• after “At Temp” test time expires, the next enabled setpoint is
cycled.
This parameter may be changed using the Encoder.
Part
Function