9330 Operating Manual
Interface Inc.
● 7401 East Butherus Drive, Scottsdale, Arizona 85260 USA ● Phone 480.948.5555 ● Fax 480.948.1924
www.interfaceforce.com
● Email:
●
800.947.5598
Document Number 15-216 Rev B
Page 30 of 74
6 Example scaling a channel using model WMC-100 load cell with 100 lbf capacity and 1.9587
mV/V output. After entering the values into the dialog box you must click “Calculate” and then
“OK/Set”.
F
IGURE
12
-
E
XAMPLE OF
C
ALIBRATION FOR A
WMC-100
L
OAD
C
ELL
Distance Offset
1. To change the distance of the origin, this setting may be access in the sensor option
2. Select the corresponding direction and the distance.
3. Can be set in meters or millimeters.
Press Calculate to scale and then OK/Set to save