2. BSC8 has a slightly different add channel box. Select Dev1 instead of Com port. Please
remember to open the needed amount of input channels.
F
IGURE
8
-
E
XAMPLE
BSC8
D
EVICE
3.
Each channel must now be scaled using the “SCALING” dialog box. Each channel must be scaled
independently. If the BSC8 was purchased with Interface load cells and a System Setup and
Scaling then the scaling values will be taken from the “Load Cell / BSC8 Digital Bridge Amplifier
Calibration Certificate”
4.
F
IGURE
9
-
E
XAMPLE OF
S
CALING
4.1. Physical full scale is typically the capacity of the sensor.
4.2. Electrical full scale output is the output of the sensor at the Physical full scale.
4.3. Input Range is always 2 mV/V and should not be changed.