![Interface 9840-400 Manual Download Page 7](http://html1.mh-extra.com/html/interface/9840-400/9840-400_manual_2076317007.webp)
I N T E R F AC E M O D E L 9 8 4 0
MO D E L 9 8 4 0 P G 7
P U B .
6 0 3 2 - 0 1
Each message will be displayed for about 3 seconds. After that the screen will start
showing the Main Menu where one of three modes (Run Mode, Setup Mode, Calibration
Mode) of operation can be selected, as shown below.
MODEL 9840 Main Menu
Run Mode will bring the unit to a screen with 4 displays depicting all 4 channels
simultaneously in their default state of Load A through Load D shown in pounds (Lb). The
example below shows Load A is 0.00116 mvV and Load B is -0.00399 mvV on the top
row. The second row of the display shows Load C is 0.00067 mvV and Load D is -
0.01952 mvV. The “T” TEDS
OPT
annunciator indicates presence or non-presence of a
TEDS
OPT
sensor. If it is on solid, a TEDS
OPT
sensor is attached to that specific channel. If
it is blinking a non TEDS
OPT
sensor is attached. The “T” is not displayed when the
TEDS
OPT
feature is turned off.
To change the Item or Unit of any of the displays select the display to be changed on the
left screen (when selected there will be a red ring around it) press the Up/Down Item
arrows on left side of the right screen to cycle through the item options active for that
channel. The rightmost Up/Down arrow buttons on the right screen will cycle through the
units that can be used. For example, with the Item “Load A” showing you would see Lb,
kg, N, PSI, MPa, Klb, kN, t, mV/V, grams, and then back to Lb. If a Torque Cell is active
the unit options would be LbI, NM, OzI, and mV/V.
#
Message
Meaning
1
Version X.Y.Z
This is the version number.
2
Serial # 12345
Option # 114236
This is the serial number.
This is the option number.
Touch Screen Vx.y.z
This is the version number for the touch screen.