I N T E R F AC E M O D E L 9 8 4 0
MO D E L 9 8 4 0 P G 1 9
P U B .
6 0 3 2 - 0 1
>>Filter Window
This option allows you to enable and set, or disable, a window to be used to turn the
filters on or off. If there is a sudden change in the measurement that exceeds this set
window, the filter is temporarily turned off to allow the displayed reading to correct itself
with the actual reading. This enables the user to see the sudden change in measured
data rather than being filtered out.
To turn a window on for a specific channel, use the
Continue
button to progress
through the different steps. First select the channel you wish to add a filter window
to, and choose to turn “On”. You will then be asked if you wish to use the
previously saved window value (0.0000 Lb or LbI default for Load/Torque cell).
“Yes” will complete the setup with the displayed value, save it and exit you out of
the filter window setup. “No” will advance you further, in order to enter a new
value. Select the unit first then the numerical value. Hitting either
Enter
button will
complete the filter window setup with the newly set value.
To change the window, simply repeat the steps above. Instead of 0 and Lbs or LbI
on the “old window” display, you will see the most recently entered value.
Note that changing a cell type on a channel will automatically disable the filter
window for that channel, reset the units to Lb or LbI, and set the window value to
0. The window must then be reset and re-enabled by the user.
>>Decimal Select
This lets you set the maximum number of digits to the right of the decimal, for each
channel, that are displayed or printed for any of the item selections. When large cells
are selected the number of decimal digits shown may be limited, for example a 100 Lb
cell will allow 3 digits, while a 1000 Lb cell will only allow 2 digits. Also, if a number is
too large to fit on the display some decimal digits may be dropped (the display will
“auto-range”).
>>Count by Select
This allows you to choose the smallest unit of counting for the display of the chosen
channel item. Counting by choices include: 1, 2, 5, 10, and 20.
Note that the operation of the analog outputs and the limits is not effected by the
settings of “decimal point” or “smallest count”. Thus if you set a limit to be activated at
10.001 Lbs, and have “smallest count” set for 20, the limit is activated at 10.001 Lbs
even though the display is only showing 10.000.
>>Select User Sums
This setting provides up to four different User Sum combinations: Usum1, Usum2,
Usum3 and Usum4. These combinations can be used as an item selection with a user
designated unit. Each User Sum is any combination of the four channels summed
together and displayed on the touchscreen, or serial terminal.
Note: These sums will only appear in the item lists if all selected channels for that Sum
are of the same cell type (Load or Torque).