I N T E R F AC E M O D E L 9 8 4 0
MO D E L 9 8 4 0 P G 4 5
P U B .
6 0 3 2 - 0 1
Example:
@123CS553.26#
Acknowledge:
@123 Calibrate Shunt Command - Reading...
[10 seconds delay]
@123 Calibrate Command - Reading for Shunt Check...
[10 seconds delay]
@123 Calibrate Command Completed
Ch C = Ch A = S/N 123456, 1000.0 Lb , 4.50014 mV/v,
10.00 V , Cal on Oct27-99, 553.26 Lb Shunt
Ch D = S/N 63220, 100.00 Lb , 2.99984 mV/v,
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v,
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt
Calibrate Counts-Per-Inch
OPT
Command (CI)
This command has two subcommands, a view command (CIV) and a set command
(CIS).
The Calibration counts per Inch View (CIV) command displays the current setting of
the encoder counts per inch constant.
Example:
@123CIV
Acknowledge:
@123 Encoder is 32000 counts per inch
The Calibration counts per Inch Set (CIS) command allows you to change the current
setting of the encoder counts per inch constant.
Example:
@123CIS64000#
Acknowledge:
@ 123 Encoder is 64000 counts per inch
OPTION ~
Limits Command (L)
This command allows you to view or set the 4 contact closure limit switches on the Model
9840. There are three subcommands Limit View (LiV), Limit Set (LiS), and Limit Reset
(LiR). Note that changing a cell type on a channel will automatically erase limits that were
set for that channel. They must then be reset by the user.
Limit View Command (LiV)
The Limit View (LiV) command is used to view the current settings of one of the limits.
The format is L(limit number)V.
Example:
@123L1V
Acknowledge:
@ 123 Lim 1 NO Enabled Load A Lb Set 20.00 Trip>Set Latch Off
Reset 1.000