4

Disassembling the Console

2.  Disconnect the contamination guard tubing 

from the vacuum port of the canister. 

3.  Open cartridge door, if necessary press the pinch valve 

button, and remove the cartridge from the console by 
sliding it up.

4.  Detach the handpiece cable from the console by 

gently pulling on the handpiece cable connector in 
the handpiece port on the front of the console.

5.  Unplug the power cord from the wall receptacle 

and the console.

6.  Disconnect the footswitch cable connector from 

the back of the console.

7.  Discard the used/contaminated items in compliance with hospital policy for contaminated waste.

Please reference the table “CUSA® Clarity Items to Keep and Discard After a Surgical Procedure” on page 2.

1.  Detach the aspiration tubing from the patient port on the 

canister.

NOTE

The contamination guard should not be removed 
from the console unless it is time to replace it 
(every six months or when the color changes).

Summary of Contents for CUSA Clarity

Page 1: ...Integra CUSA Clarity Ultrasonic Surgical Aspirator System Cleaning Instruction Guide...

Page 2: ...the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed It is intended as a guide for using the CUSA...

Page 3: ...the irrigation tubing from the clip at the base of the handpiece Disassembling the System CUSA Clarity Items to Keep and Discard After a Surgical Procedure Keep Discard Handpiece Nosecone Contaminati...

Page 4: ...nch color side towards handpiece over the tip 10 Rotate the torque wrench counter clockwise 11 Unthread the tip from the handpiece connecting body by turning the tip counter clockwise by hand 12 Disca...

Page 5: ...of the console 5 Unplug the power cord from the wall receptacle and the console 6 Disconnect the footswitch cable connector from the back of the console 7 Discard the used contaminated items in compli...

Page 6: ...hat the four anti static wheels function correctly CAUTION Product damage may result if you do not follow the instructions described in the cleaning section WARNING Electric Shock Hazard Always unplug...

Page 7: ...ning brush is provided non sterile and is not intended for use in a sterile environment It is intended to be used and stored only in the reprocessing unit of the healthcare facility Automated Cleaning...

Page 8: ...nnect the footswitch from the console 2 Wipe it clean 3 If the footswitch is contaminated with blood or fluid a Immerse the footswitch in warm water containing detergent or disinfectant b Rinse the fo...

Page 9: ...nded to protect the CUSA Clarity Handpiece EUT and Torque Base during steam sterilization and during transfer to the sterile field The handpiece and components must be cleaned prior to sterilization s...

Page 10: ...lot 2 Wrap the handpiece cable around the posts 3 Place the handpiece connector in the appropriate slot 4 Optional Place the nosecone in the appropriate slot 5 Optional Place the flue anywhere on the...

Page 11: ...Prevac 4 min None Sterilization Cycles for the 23 kHz and 36 kHz Handpiece and Components CAUTION Do not sterilize flues with temperatures in excess of 134 C 273 2 F as this may damage the flue The C...

Page 12: ...emark of Integra LifeSciences Corporation or its subsidiaries Sani Cloth is a registered trademark of Professional Disposables International Inc neodisher is a trademark of Chemische Fabrik Dr Weigert...