Cleaning the Handpiece, EUT, Sterilization Tray, and Torque Base

The handpieces, extended use tips, sterilization tray, and torque base may be cleaned by either the manual cleaning or autowash 
cycle process prior to sterilization. For Sterilization Parameters, see “

Sterilizing the System

” on page 8.

Manual Cleaning

Use the provided cleaning brush to clean out the internal channels of both the handpiece and tip. Continue to pass the brush 
through the tip and handpiece until it comes out clean.

1.  Clean the exterior of the handpiece and tip with a lint free cloth soaked with warm water and a non-alkaline detergent. 

2.  Use the cloth to manually clean the junction where the tip attaches.

3.  Rinse the entire handpiece and tip thoroughly with water and dry with a soft cloth.

Cleaning Brush

The CUSA Clarity Cleaning Brush is intended to be used primarily to clean the internal pathways, as needed, of the CUSA Clarity 
handpiece following use, during the manual cleaning process or as the first step of  the automated cleaning process, just prior to 
autowashing. 

The CUSA Clarity cleaning brush is provided non-sterile and is not intended for use in a sterile environment. It is intended to be used 
and stored only in the reprocessing unit of the healthcare facility. 

Automated Cleaning:

Before automated cleaning, use the provided cleaning brush, dipped in cold tap water, to thoroughly clean the internal channels of 
the handpiece and tip. Continue to pass the brush through the tip and handpiece until it comes out clean. Rinse the exterior and in-
terior of the tip in cold running tap water for 30 seconds. For initial exterior cleaning of each component, prepare a cleaning solution 
using a neutral or mid-alkaline detergent in accordance with the manufacturer’s instructions. Using lint-free cloths soaked in the 
prepared cleaning solution, wipe clean each component.

•  In a Steelco DS50 DRS or equivalent washer /disinfector: 
 

•   Perform a prewash cycle of 5 minutes at 43°C. 

 

•    Perform a cleaning cycle using a mid-alkaline pH detergent  

(pH 10.0 – pH 11.5) for 10 minutes at 60°C. Prepare the detergent solution according to the manufacturer’s recommenda-

tions.

 

•   Perform a disinfection cycle for 10 minutes at 93°C.

 

•   Dry for 20 minutes at 55°C.

NOTE

Do not clean the sterilizer tray with abrasives. Product damage will result.

NOTE

Do not submerge the handpiece cable connector.

Summary of Contents for CUSA Clarity

Page 1: ...Integra CUSA Clarity Ultrasonic Surgical Aspirator System Cleaning Instruction Guide...

Page 2: ...the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed It is intended as a guide for using the CUSA...

Page 3: ...the irrigation tubing from the clip at the base of the handpiece Disassembling the System CUSA Clarity Items to Keep and Discard After a Surgical Procedure Keep Discard Handpiece Nosecone Contaminati...

Page 4: ...nch color side towards handpiece over the tip 10 Rotate the torque wrench counter clockwise 11 Unthread the tip from the handpiece connecting body by turning the tip counter clockwise by hand 12 Disca...

Page 5: ...of the console 5 Unplug the power cord from the wall receptacle and the console 6 Disconnect the footswitch cable connector from the back of the console 7 Discard the used contaminated items in compli...

Page 6: ...hat the four anti static wheels function correctly CAUTION Product damage may result if you do not follow the instructions described in the cleaning section WARNING Electric Shock Hazard Always unplug...

Page 7: ...ning brush is provided non sterile and is not intended for use in a sterile environment It is intended to be used and stored only in the reprocessing unit of the healthcare facility Automated Cleaning...

Page 8: ...nnect the footswitch from the console 2 Wipe it clean 3 If the footswitch is contaminated with blood or fluid a Immerse the footswitch in warm water containing detergent or disinfectant b Rinse the fo...

Page 9: ...nded to protect the CUSA Clarity Handpiece EUT and Torque Base during steam sterilization and during transfer to the sterile field The handpiece and components must be cleaned prior to sterilization s...

Page 10: ...lot 2 Wrap the handpiece cable around the posts 3 Place the handpiece connector in the appropriate slot 4 Optional Place the nosecone in the appropriate slot 5 Optional Place the flue anywhere on the...

Page 11: ...Prevac 4 min None Sterilization Cycles for the 23 kHz and 36 kHz Handpiece and Components CAUTION Do not sterilize flues with temperatures in excess of 134 C 273 2 F as this may damage the flue The C...

Page 12: ...emark of Integra LifeSciences Corporation or its subsidiaries Sani Cloth is a registered trademark of Professional Disposables International Inc neodisher is a trademark of Chemische Fabrik Dr Weigert...

Reviews: