5
Integra
®
CUSA
®
Clarity – Ultrasonic Surgical Aspirator System
Cleaning Instruction Guide
Detaching the Contamination Guard
Powering Down the System
Press the
Power
button on the front of the console. The touchscreen will go dark. Do not unplug power cord until system
has powered down.
Cleaning the System
Cleaning the Console, Touchscreen, and Cart surfaces
1. Unplug the power cord from the wall receptacle and the console.
2. Clean the surface with a lint-free wipe soaked with one of the following:
• 70% IPA
• Super Sani-Cloth®
• Sani-Cloth® Bleach
3. Make sure the surfaces remain visibly wet for a minimum of two (2) minutes, then air dry.
4. Clean the wheels at the base of the system cart to ensure that the four anti-static wheels function correctly.
CAUTION
Product damage may result if you do not follow the
instructions described in the cleaning section.
WARNING
Electric Shock Hazard
- Always unplug
the CUSA Clarity System before cleaning it.
NOTE
Never spray any cleaner directly onto
the touch screen.
NOTE
The contamination guard should not be removed from the console unless
it is time to replace it (every six months or when the color changes).
NOTE
Make sure that the surface of the wheels is free
from dirt and dust. Make sure that the surface is
completely dry before using the system again.
WARNING
Do not allow fluids to enter the console.
1. Detach the handpiece cable from the console by
gently pulling on the handpiece cable connector
in the handpiece port on the front of the console.
2. Discard the contamination guard in compliance
with hospital policy for contaminated waste.