background image

Integra

®

 CUSA

®

 Clarity – Ultrasonic Surgical Aspirator System

Cleaning Instruction Guide

USA 800-654-2873 

 

n

  

888-980-7742 fax

Interna1 609-936-5400

  

n

  

+

609-750-4259 fax

integralife.com/contact

United States, Canada, Asia, Pacific, Latin America

Availability of these products might vary from a given country or region to another, as a result of specific local regulatory approval or clearance requirements for sale in such country or region. 

 

n

 Non contractual document. The manufacturer reserves the right, without prior notice, to modify the products in order to improve their quality.

n

 Warning: Applicable laws restrict these products to sale by or on the order of a physician.

n

 Consult product labels and inserts for any indication, contraindications, hazards, warnings, precautions, and instructions for use.

For more information or to place an order, please contact:

Integra, the Integra logo, and CUSA are registered trademarks of Integra LifeSciences Corporation or its subsidiaries in the United States and/or other countries. CUSA Quick Connect 
is a trademark of Integra LifeSciences Corporation or its subsidiaries. Sani-Cloth is a registered trademark of Professional Disposables International, Inc. neodisher is a trademark 
of Chemische Fabrik Dr. Weigert GmbH & Co. KG. All other trademarks and trade names are the property of their respective owners. ©2019 Integra LifeSciences Corporation. 
All rights reserved. Printed in USA. 0M           0932889-3-EN

Integra LifeSciences (Ireland) Limited
IDA Business &Technology Park
Sragh
Tullamore
County Offaly
Ireland

Manufacturer:

Sterilizing the Handpiece and Components with Steam

Sterilizing the handpiece with steam depends on the following factors:

• Temperature
• Exposure time
• Population and resistance of resident bioburden
• Method of air removal from the autoclave

Use the validated steam sterilization cycle parameters in these instructions. If you deviate from this recommended method of steril-
izing, it is your Health Care Facility’s responsibility to validate the deviation. For sterilizing the handpiece with steam, do not exceed 
137°C (278.6°F).

CAUTION

Do not use sterilization methods involving temperatures in 
excess of 137ºC (278.6°F) as this may damage the handpiece.

CAUTION

Do not sterilize flues with temperatures in excess of 134°C 
(273.2°F) as this may damage the flue.

Summary of Contents for CUSA Clarity

Page 1: ...Integra CUSA Clarity Ultrasonic Surgical Aspirator System Cleaning Instruction Guide...

Page 2: ...the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed It is intended as a guide for using the CUSA...

Page 3: ...the irrigation tubing from the clip at the base of the handpiece Disassembling the System CUSA Clarity Items to Keep and Discard After a Surgical Procedure Keep Discard Handpiece Nosecone Contaminati...

Page 4: ...nch color side towards handpiece over the tip 10 Rotate the torque wrench counter clockwise 11 Unthread the tip from the handpiece connecting body by turning the tip counter clockwise by hand 12 Disca...

Page 5: ...of the console 5 Unplug the power cord from the wall receptacle and the console 6 Disconnect the footswitch cable connector from the back of the console 7 Discard the used contaminated items in compli...

Page 6: ...hat the four anti static wheels function correctly CAUTION Product damage may result if you do not follow the instructions described in the cleaning section WARNING Electric Shock Hazard Always unplug...

Page 7: ...ning brush is provided non sterile and is not intended for use in a sterile environment It is intended to be used and stored only in the reprocessing unit of the healthcare facility Automated Cleaning...

Page 8: ...nnect the footswitch from the console 2 Wipe it clean 3 If the footswitch is contaminated with blood or fluid a Immerse the footswitch in warm water containing detergent or disinfectant b Rinse the fo...

Page 9: ...nded to protect the CUSA Clarity Handpiece EUT and Torque Base during steam sterilization and during transfer to the sterile field The handpiece and components must be cleaned prior to sterilization s...

Page 10: ...lot 2 Wrap the handpiece cable around the posts 3 Place the handpiece connector in the appropriate slot 4 Optional Place the nosecone in the appropriate slot 5 Optional Place the flue anywhere on the...

Page 11: ...Prevac 4 min None Sterilization Cycles for the 23 kHz and 36 kHz Handpiece and Components CAUTION Do not sterilize flues with temperatures in excess of 134 C 273 2 F as this may damage the flue The C...

Page 12: ...emark of Integra LifeSciences Corporation or its subsidiaries Sani Cloth is a registered trademark of Professional Disposables International Inc neodisher is a trademark of Chemische Fabrik Dr Weigert...

Reviews: