Integra
®
CUSA
®
Clarity – Ultrasonic Surgical Aspirator System
Cleaning Instruction Guide
USA 800-654-2873
n
888-980-7742 fax
Interna1 609-936-5400
n
+
1
609-750-4259 fax
integralife.com/contact
United States, Canada, Asia, Pacific, Latin America
Availability of these products might vary from a given country or region to another, as a result of specific local regulatory approval or clearance requirements for sale in such country or region.
n
Non contractual document. The manufacturer reserves the right, without prior notice, to modify the products in order to improve their quality.
n
Warning: Applicable laws restrict these products to sale by or on the order of a physician.
n
Consult product labels and inserts for any indication, contraindications, hazards, warnings, precautions, and instructions for use.
For more information or to place an order, please contact:
Integra, the Integra logo, and CUSA are registered trademarks of Integra LifeSciences Corporation or its subsidiaries in the United States and/or other countries. CUSA Quick Connect
is a trademark of Integra LifeSciences Corporation or its subsidiaries. Sani-Cloth is a registered trademark of Professional Disposables International, Inc. neodisher is a trademark
of Chemische Fabrik Dr. Weigert GmbH & Co. KG. All other trademarks and trade names are the property of their respective owners. ©2019 Integra LifeSciences Corporation.
All rights reserved. Printed in USA. 0M 0932889-3-EN
Integra LifeSciences (Ireland) Limited
IDA Business &Technology Park
Sragh
Tullamore
County Offaly
Ireland
Manufacturer:
Sterilizing the Handpiece and Components with Steam
Sterilizing the handpiece with steam depends on the following factors:
• Temperature
• Exposure time
• Population and resistance of resident bioburden
• Method of air removal from the autoclave
Use the validated steam sterilization cycle parameters in these instructions. If you deviate from this recommended method of steril-
izing, it is your Health Care Facility’s responsibility to validate the deviation. For sterilizing the handpiece with steam, do not exceed
137°C (278.6°F).
CAUTION
Do not use sterilization methods involving temperatures in
excess of 137ºC (278.6°F) as this may damage the handpiece.
CAUTION
Do not sterilize flues with temperatures in excess of 134°C
(273.2°F) as this may damage the flue.