3
Integra
®
CUSA
®
Clarity – Ultrasonic Surgical Aspirator System
Cleaning Instruction Guide
5. Slide the flue off of the nosecone base along the tip.
6. Twist off the nosecone counter-clockwise until it unlocks.
7. Slide the nosecone off the tip.
8. Place the handpiece on the torque base as shown and make
sure that the 2 flat metal sides fit snugly in the metal slot at
the end of the base.
9. Hold the handpiece in the base and place the torque
wrench (color side towards handpiece) over the tip.
10. Rotate the torque wrench counter-clockwise.
11. Unthread the tip from the handpiece connecting
body by turning the tip counter-clockwise by hand.
12. Discard the items (listed in the table in “
CUSA® Clarity
Items to Keep and Discard After a Surgical Procedure
”
on page 2) in compliance with hospital policy for con-
taminated waste.
CAUTION
To avoid product damage, always use the torque base
to hold the handpiece while using the torque wrench to
tighten or loosen the tip. Avoid over torqueing the tip.
23 kHz
36 kHz
23 kHz
36 kHz
23 kHz
36 kHz
23 kHz
36 kHz
23 kHz
36 kHz
23 kHz
36 kHz
23 kHz
36 kHz