139
Understanding PDM and Pod Function 12
12 Understanding PDM and Pod Function
The calculations for the 50% increase temp basal in the preceding figure are:
Segment
boundaries*
Basal rate of
Basal Program
(U/hr)
50% increase
(U/hr)
Resulting temp
basal rate:
(U/hr)
Midnight–7:00
0.20
7:00–10:00
0.20
0.20 x 50% = 0.10
0.20 + 0.10 = 0.30
10:00–14:00
0.60
0.60 x 50% = 0.30
0.60 + 0.30 = 0.90
14:00–16:30
0.30
0.30 x 50% = 0.15
0.30 + 0.15 = 0.45
16:30 – midnight
0.30
* Segments are defined by the currently scheduled Basal Program.
Temp basal from 07:00 to 16:30
Basal Program
Actual insulin delivered
midnight
midnight
Noon
0
0.2
0.4
0.6
0.8
1.0
midnight
Noon
midnight
Basal R
at
e (U/hr)
50% decrease
50% increase
Temp basal set to %
midnight
midnight
Noon
0
0.2
0.4
0.6
0.8
1.0
midnight
Noon
midnight
Basal R
at
e (U/hr)
set to 0.1 U/hr
set to 0.9 U/hr
Temp basal set to U/hr
Effective Date: 27SEP2019, ECO-000240
Summary of Contents for INT2-D001-MM-OUS2
Page 11: ...x Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 17: ...xvi Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 37: ...20 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 53: ...36 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 73: ...56 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 85: ...68 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 141: ...124 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 177: ...160 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 211: ...194 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 221: ...My Notes Effective Date 27SEP2019 ECO 000240...