123
Alarms, Notifications, Communication Errors 10
10 Alarms, Notifications, Communication Errors
Silencing an Alarm
You can usually silence Pod or PDM alarms by tapping a button on the alarm
screen. If the alarm continues, or if you are discarding a Pod and want to prevent a
future alarm, follow the directions in this section.
Pod alarm
To permanently silence a Pod alarm:
1. If the Pod is on your body, remove it.
2. Peel back a little bit of the adhesive
pad from the bottom of the Pod at the
square end (see figure).
3. Locate the alarm shut-off port to the
right of the gold circle. The alarm shut-
off port can be felt with a fingernail or
paper clip as a soft plastic.
4. Firmly press a paper clip or similar
item straight down into the alarm shut-
off port. If an alarm is sounding, the
alarm stops. You need to apply enough
force to break a thin layer of plastic.
PDM alarm
If a PDM alarm is not silenced when you tap the alarm screen’s button:
1. Press and hold the Power button, then tap
Power off
.
2. Take out and re-insert the PDM’s battery. For instructions on removing and
replacing the PDM battery, see “Inserting or Replacing the PDM Battery” on
page 131.
3. Press and hold the Power button to turn the PDM back on.
This action silences the PDM’s alarm. Your history records and personal settings
are preserved. However, you may have to reset the time and date and change
your Pod. After you reset the time, the Bolus Calculator is temporarily disabled.
A PDM message tells you when the Bolus Calculator can be used again. You can
deliver a manually calculated bolus when the Bolus Calculator is disabled.
Fill port
Gold circle
Alarm shut-off
port
Press down
firmly
.
Effective Date: 27SEP2019, ECO-000240
Summary of Contents for INT2-D001-MM-OUS2
Page 11: ...x Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 17: ...xvi Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 37: ...20 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 53: ...36 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 73: ...56 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 85: ...68 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 141: ...124 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 177: ...160 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 211: ...194 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Page 221: ...My Notes Effective Date 27SEP2019 ECO 000240...