Preliminary Pages
6
M16-16905-EN
Hazard - Set the appropriate limits before performing loop tuning or running
waveforms or tests.
Operational limits are included within your testing system to suspend motion or
shut off the system when upper and/or lower bounds of actuator or crosshead
travel, or force or strain, are reached during testing. Correct setting of
operational limits by the operator, prior to testing, will reduce the risk of damage
to test article and system and associated hazard to the operator.
Electrical Hazard - Disconnect the electrical power supply before removing
the covers to electrical equipment.
Disconnect equipment from the electrical power supply before removing any
electrical safety covers or replacing fuses. Do not reconnect the power source
while the covers are removed. Refit covers as soon as possible.
Rotating Machinery Hazard - Disconnect power supplies before removing
the covers to rotating machinery.
Disconnect equipment from all power supplies before removing any cover which
gives access to rotating machinery. Do not reconnect any power supply while
the covers are removed unless you are specifically instructed to do so in the
manual. If the equipment needs to be operated to perform maintenance tasks
with the covers removed, ensure that all loose clothing, long hair, etc. is tied
back. Refit covers as soon as possible.
Hazard - Shut down the hydraulic power supply and discharge hydraulic
pressure before disconnection of any hydraulic fluid coupling.
Do not disconnect any hydraulic coupling without first shutting down the
hydraulic power supply and discharging stored pressure to zero. Tie down or
otherwise secure all pressurized hoses to prevent movement during system
operation and to prevent the hose from whipping about in the event of a
rupture.
Hazard - Shut off the supply of compressed gas and discharge residual gas
pressure before you disconnect any compressed gas coupling.
Do not release gas connections without first disconnecting the gas supply and
discharging any residual pressure to zero.
Warnings
Summary of Contents for AVE2
Page 1: ...Instron Video Extensometer AVE2 and SVE2 Operator s Guide M16 16905 EN Revision D...
Page 8: ...Preliminary Pages 8 M16 16905 EN...
Page 12: ...Preliminary Pages 12 M16 16905 EN...
Page 16: ...Chapter Introduction 16 M16 16905 EN...
Page 24: ...Chapter Specifications 24 M16 16905 EN...
Page 34: ...Chapter Installation 34 M16 16905 EN Figure 9 AVE2 on a 5940 single column testing machine A...
Page 46: ...Chapter Installation 46 M16 16905 EN Figure 21 AVE2 on an 8801 series testing machine...
Page 50: ...Chapter Installation 50 M16 16905 EN...
Page 76: ...Chapter Set up other software not Bluehill 76 M16 16905 EN...
Page 84: ...Chapter Maintenance 84 M16 16905 EN...
Page 87: ......
Page 88: ...www instron com...