Chapter: Mark specimens
54
M16-16905-EN
•
wicking can occur when testing woven specimens and the wet ink is drawn, by
capillary action, into the fibers of the specimen.
If bleeding or wicking is a consistent problem, you should consider using tape dots.
•
Loss of detection of ink dots during tests can be caused by:
•
change of specimen color during testing so that contrast between the dot and
the specimen is lost - consider using a different ink color
•
fading of the dot due to excessive extension during the test - consider using tape
dots
•
flaking of the ink from the specimen - make sure the ink is compatible with the
specimen material and consider testing the specimen while the ink is still wet
•
the dot moves outside the field of view of the lens because specimen extension
was larger than expected - use a lens with a larger field of view
Summary of Contents for AVE2
Page 1: ...Instron Video Extensometer AVE2 and SVE2 Operator s Guide M16 16905 EN Revision D...
Page 8: ...Preliminary Pages 8 M16 16905 EN...
Page 12: ...Preliminary Pages 12 M16 16905 EN...
Page 16: ...Chapter Introduction 16 M16 16905 EN...
Page 24: ...Chapter Specifications 24 M16 16905 EN...
Page 34: ...Chapter Installation 34 M16 16905 EN Figure 9 AVE2 on a 5940 single column testing machine A...
Page 46: ...Chapter Installation 46 M16 16905 EN Figure 21 AVE2 on an 8801 series testing machine...
Page 50: ...Chapter Installation 50 M16 16905 EN...
Page 76: ...Chapter Set up other software not Bluehill 76 M16 16905 EN...
Page 84: ...Chapter Maintenance 84 M16 16905 EN...
Page 87: ......
Page 88: ...www instron com...