Chapter: Set up Bluehill Universal
®
software
66
M16-16905-EN
8. Follow the instructions in the calibration dialog to select and focus the lens.
a Select the lens that you have installed from the drop-down list.
b Focus of the lens until the check mark appears, then lock the thumb screw.
9. Replace the red lens cover (AVE2 only).
10. In the image window in the Calibration screen, drag the dotted center line so that it
is in the center of the dot matrix pattern.
11. Draw a box around the area of the dot matrix that you want to calibrate. The border
of the box displays green.
If you are using the 6mm lens, draw the box so that it includes both calibration bars.
12. Select Calibrate.
13. When calibration is complete, the calibration status updates and a yellow box
appears indicating the region in the field of view that generates the highest accuracy.
You should ensure that the dots defining the gauge length on each specimen are
within this yellow box at the start of each test.
14. Select AVE2 (or SVE2).
The dialog moves to the AVE2 (or SVE2) screen.
15. Remove the calibration bar from the grips.
16. If you are using the video extensometer on an 8800 or ElectroPuls testing system
with an 8800T or 8800D controller, you now need to manually calibrate the exten-
someter in 8800/ElectroPuls console. As part of the manual calibration, set the
transducer type to Strain (Calibration wizard > Transducer details screen). For more
details refer to “Using the AVE2 with an 8800T or 8800D Controller” in the Console
online help.
The strain transducers for the video extensometer are now calibrated.
Summary of Contents for AVE2
Page 1: ...Instron Video Extensometer AVE2 and SVE2 Operator s Guide M16 16905 EN Revision D...
Page 8: ...Preliminary Pages 8 M16 16905 EN...
Page 12: ...Preliminary Pages 12 M16 16905 EN...
Page 16: ...Chapter Introduction 16 M16 16905 EN...
Page 24: ...Chapter Specifications 24 M16 16905 EN...
Page 34: ...Chapter Installation 34 M16 16905 EN Figure 9 AVE2 on a 5940 single column testing machine A...
Page 46: ...Chapter Installation 46 M16 16905 EN Figure 21 AVE2 on an 8801 series testing machine...
Page 50: ...Chapter Installation 50 M16 16905 EN...
Page 76: ...Chapter Set up other software not Bluehill 76 M16 16905 EN...
Page 84: ...Chapter Maintenance 84 M16 16905 EN...
Page 87: ......
Page 88: ...www instron com...