insize ISR-S300 Operation Manual Download Page 9

15

16

Note:  Measurements  should  always  be  made  with  the 
temperature  of  the  room,  of  the  part  and  of  the  instrument 
constant and, where possible, at 20°C.

Table 1. Relationship between roughness and cut-off.

Calibration

You are recommended not to use the specimen plate more than is 
strictly  necessary,  to  avoid  deterioration  caused  by  possible 
scoring of the stylus which could falsify the roughness values.

Recharging And Replacing The Accumulator

When the accumulator charge is low, the instrument signals    on 
the  extreme  right  of  the  first  line  of  the  display.  The  instrument 
should  then  be  connected,  as  quickly  as  possible,  to  the  power 
supply  provided;  then  plug  the  power  supply  into  a  mains  outlet, 
after making sure that the voltage is that required (220V 50/60Hz). 
The power supply plug should be inserted in the socket shown in 
figure  2,  on  the  rear  of  the  ISR-S400,  and  making  sure  that  it  is 
plugged in fully; leave to charge for at least 3 hours. A full charge 
takes  4  hours.  No  problem  will  arise  if  the  power  supply  is  left 
connected for more than this time, because it is provided with an 
automatic charge control means.
It  is  possible  to  use  the  instrument  even  while  the  accumulator 
battery is recharging.

Note:  do  not  store  the  instrument  for  long  periods  with  the 
accumulator battery discharged, as it will become damaged. 
If  left  idle  for  long  periods,  check  the  charge  level  at  least 
once  every  30  days.  If  the  accumulator  discharges  too 
deeply, it will not be possible to switch the roughness tester 
on.  Do  not  replace  the  accumulator  with  zinc-carbon  or 
alkaline  batteries.  These  batteries  can  explode  if  you  try  to 
recharge  them  and,  in  any  case,  may  cause  the  instrument 
damage when their charge is low.

1

2

T h e  a c c u m u l a t o r  c a n  w i t h s t a n d  a  c e r t a i n  n u m b e r  o f 
charge/discharge  cycles,  beyond  which  it  can  no  longer  be 
charged  and  must  be  replaced.  Use  an  8.4V-170mAh  NiCd  or 
NiMH type accumulator, PP3 format.
Remove  the  six  base  fixing  screw  of  the  instrument  cover  then 
pull  the  cover  from  the  top,

 

detach  the  accumulator  from  the 

button  connection  and  insert  the  new  one.  Make  sure  that  the 
accumulator  does  not  touch  the  contacts  with  the  polarities 
inverted!
Insert again the cover and fix it with the screws to the instrument 
base.

Summary of Contents for ISR-S300

Page 1: ...MN ISR 300 S E www insize com ROUGHNESS TESTER OPERATION MANUAL ISR S300 V0 ...

Page 2: ... a microprocessor it uses an alphanumeric liquid crystal display LCD and the accumulator adopted permits considerable operating autonomy All the messages and menus can be displayed in one of the following five selectable languages Italian English French German Spanish and Portuguese The instrument consists of the electronic unit the miniaturised traverse unit incorporated and where applicable a pr...

Page 3: ...n of the items on the top part is as follows 1 2 On off START start measurement STOP stop measurement key 3 Key for displaying the result of the measurement displaying and presetting the cut off and cut off number If pressed when in Liquid crystal display LCD the main menu causes the higher level menu to return on display 4 Scroll key If pressed when the result of a measurement is on display bring...

Page 4: ... shows 1 The results of a measurement 2 The cut off 3 The cut off number 4 An error message To exit from the main menu and the submenus press the and the first one immediately starts the measure the second in the submenus makes to climb a level in the main menu it makes to visualize the parameters The structure of the menu can be seen in the figure that follows The selections made on the submenus ...

Page 5: ...ayed because the assessment of the parameters is strictly related to λc and to the number of λc used When the cut off or cut off number is changed on an instrument already preset for measurement setting of the cut off may be terminated with the key which immediately starts the measuring cycle 4 How to preset the traverse length with ISO 12085 When the ISO 12085 standard is used the measurement ran...

Page 6: ...selection on the item Position and press The following indication will be obtained where the vertical arrow indicates current position of the stylus while the horizontal line below it represents the total excursion of the stylus during the positioning session In order to obtain correct measurements the arrow must be in the centre of the field defined on the display coinciding with the zero point o...

Page 7: ...rated in the examples that follow 11 12 Note if you need to check the value of a disabled parameter and you have just finished measuring simply enable the parameter concerned It is not necessary to complete the measurement The parameter Rmr represents the bearing ratio and it is also graphically printed as curve of Abbott 8 Setting the calculating standard 1 In the main menu make your selection on...

Page 8: ...menu Files Eliminating all the measurements stored 1 In the main menu make your selection on the item Files and press to bring up the submenu 2 Make your selection on the item Reset and press Press to go back to the main menu 16 Operations When performing measurements to get a correct result there are a number of hints that should be observed 1 Where possible place the instrument and the part bein...

Page 9: ... is left connected for more than this time because it is provided with an automatic charge control means It is possible to use the instrument even while the accumulator battery is recharging Note do not store the instrument for long periods with the accumulator battery discharged as it will become damaged If left idle for long periods check the charge level at least once every 30 days If the accum...

Page 10: ... receiver is not programmed correctly c The connection cable is damaged Signalled when the diagnostic checks reveal a memory defect Remedies call the technical service department Signalled at the end of a measurement when with the ISO 12085 standard selected the instrument s calculating capacity is exceeded Probable cause The profile examined cannot be gauged Resetting The Storage Memory If there ...

Reviews: