Error Messages

Probable cause:
The  instrument’s  measuring  capacity  has  been  exceeded  (Rt>300 

μ

m).

Remedies:  clean  the  part  thoroughly  and  check  that  there  are  no 
furrows at the measuring point. Try measuring a different point.

Data

 

present

 

in

 

memory

 

can

 

no

 

longer

 

be

 

used

 

to

 

gauge

 

roughness

.

Signalled

 

in

 

the

 

following

 

cases

after

 

pressing

       :

a

After

 

changing

 

the

 

cut

-

off

.

b

After

 

changing

 

the

 

cut

-

off

 

number

.

c

After

 

calibration

.

d

After

 

a

 “

measure

 

test

” 

in

 

the

 

Position function

.

Remedies

make

 

a

 

measurement

.

The value measured deviates too much from that of the specimen.
Remedies:  repeat  the  calibration  and  check  the  value  that  you  have 
preset.

The calibration data have accidentally been erased or damaged.
Signalled after pressing      .

Signalled  when  the  data  are  required  for  printing  or  displaying  or 
when  the  roughness  tester  has  to  receive  external  commands.  The 
report is generated by the instrument timing-out for 2 seconds.
Probable causes:
a) The instrument that has to receive the data is not connected or is 
switched off.
b) The receiver is not programmed correctly.
c) The connection cable is damaged.

Signalled when the diagnostic checks reveal a memory defect.
Remedies: call the technical service department.

Signalled  at  the  end  of  a  measurement  when,  with  the  ISO  12085 
s ta n d a r d  s e l e c te d ,  th e  i n s tr u me n t’s  c a l c u l a ti n g  c a p a ci ty  i s 
exceeded.
Probable cause: The profile examined cannot be gauged.

Resetting The Storage Memory

If

 

there

 

are

 

problems

 

with

 

the

 

functioning

 

of

 

the

 

menus

which

 

may 

occur

 

due

 

to

 

excessive  noise

 

when

 

the

 

roughness

 

tester

 

is

 

powered

 

from

 

the

 

mains

the

 

storage

 

memory

 

must

 

be

 

reset in

 

order

 

to

 

restore

 

proper

 

working

 

conditions

.

However

note

 

that

 

the

 

resetting

 

involves

 

the

 

elimination

 

of

 

all

 

the

 

presettings

 

and

 

flags that

 

enable

 

the

 

software

 

to

 

restore

 

the

 

condition

 

prior

 

to

 

the

 

switch

-

off

.

The

 

reset

 

is

 

performed

 

by

 

holding

 

down

 

the

 

key   

   

while switching the 

instrument on

and releasing it only after the mention appears          

                  on the display.
It  has  been  done  correctly  if,  when  you  press  the  key        again,  the 
following signal              is displayed and if, when you check in the 
menus,  all  the  functions  (print  parameters,  tolerances,  etc.)  are 
disabled, and the cut-off is equal to 0.8.
If this is not the case, leave it to switch off and repeat the operation.

Summary of Contents for ISR-S300

Page 1: ...MN ISR 300 S E www insize com ROUGHNESS TESTER OPERATION MANUAL ISR S300 V0 ...

Page 2: ... a microprocessor it uses an alphanumeric liquid crystal display LCD and the accumulator adopted permits considerable operating autonomy All the messages and menus can be displayed in one of the following five selectable languages Italian English French German Spanish and Portuguese The instrument consists of the electronic unit the miniaturised traverse unit incorporated and where applicable a pr...

Page 3: ...n of the items on the top part is as follows 1 2 On off START start measurement STOP stop measurement key 3 Key for displaying the result of the measurement displaying and presetting the cut off and cut off number If pressed when in Liquid crystal display LCD the main menu causes the higher level menu to return on display 4 Scroll key If pressed when the result of a measurement is on display bring...

Page 4: ... shows 1 The results of a measurement 2 The cut off 3 The cut off number 4 An error message To exit from the main menu and the submenus press the and the first one immediately starts the measure the second in the submenus makes to climb a level in the main menu it makes to visualize the parameters The structure of the menu can be seen in the figure that follows The selections made on the submenus ...

Page 5: ...ayed because the assessment of the parameters is strictly related to λc and to the number of λc used When the cut off or cut off number is changed on an instrument already preset for measurement setting of the cut off may be terminated with the key which immediately starts the measuring cycle 4 How to preset the traverse length with ISO 12085 When the ISO 12085 standard is used the measurement ran...

Page 6: ...selection on the item Position and press The following indication will be obtained where the vertical arrow indicates current position of the stylus while the horizontal line below it represents the total excursion of the stylus during the positioning session In order to obtain correct measurements the arrow must be in the centre of the field defined on the display coinciding with the zero point o...

Page 7: ...rated in the examples that follow 11 12 Note if you need to check the value of a disabled parameter and you have just finished measuring simply enable the parameter concerned It is not necessary to complete the measurement The parameter Rmr represents the bearing ratio and it is also graphically printed as curve of Abbott 8 Setting the calculating standard 1 In the main menu make your selection on...

Page 8: ...menu Files Eliminating all the measurements stored 1 In the main menu make your selection on the item Files and press to bring up the submenu 2 Make your selection on the item Reset and press Press to go back to the main menu 16 Operations When performing measurements to get a correct result there are a number of hints that should be observed 1 Where possible place the instrument and the part bein...

Page 9: ... is left connected for more than this time because it is provided with an automatic charge control means It is possible to use the instrument even while the accumulator battery is recharging Note do not store the instrument for long periods with the accumulator battery discharged as it will become damaged If left idle for long periods check the charge level at least once every 30 days If the accum...

Page 10: ... receiver is not programmed correctly c The connection cable is damaged Signalled when the diagnostic checks reveal a memory defect Remedies call the technical service department Signalled at the end of a measurement when with the ISO 12085 standard selected the instrument s calculating capacity is exceeded Probable cause The profile examined cannot be gauged Resetting The Storage Memory If there ...

Reviews: