Record Management
Click on the corresponding record to see the details of the record.
12
Software Information
Instruments software and hardware information can help users easily
upgrade and maintain the product, unique serial number of the
instrument software information items are displayed.
13
Parameter calibration
Before measuring, the instrument usually required calibration on
standard calibration block. The instrument is configured with a standard
calibration block, before measurement, instruments should test on the
block. Under normal circumstances, when the difference between the
measured value and the block value is in the acceptable range, the
measurement value is valid, can be measured directly.
If the measured value and the block value which difference is greater
than a accuracy error range of the instrument, or the user require high
accuracy, can be used to correct the indication calibration function and
improve measurement accuracy. The value of the calibration procedure
is as shown.
Illustration is based on a model calibrated 1.63μm steps to calibrate the
model for the actual calibration of the nominal value of the set value.
1. Under normal circumstances, the instrument in the factory have been
rigorously tested, showing error is much less than ± 10%, in this case,
the user is not showing the value of the calibration frequently used
functions.
2. After setting the calibration value, you must press the START key for
a full measurement, instrument calibration to be valid.
3. New parameters after calibration must be carried out once a
complete measurement and press the Save&Exit key is stored in the
instrument.
4. Press "Exit"key to return the menu without saving calibration results.
14
Print Function setup
The instrument can be tested according to the actual requirements of
any parameter selection Print or Print All, the steps shown in Figure.