insize ISR-C300 Operation Manual Download Page 4

5

6

 

 

 

When battery voltage is too low (that is, battery voltage symbol display

   

       

   

       on  screen  to  prompt  low  voltage),  the  instrument  should  be 
charged as   soon as possible. USB port of the instrument for charging. 
You can use the built-in power adapter for charging, you can also use 
computer's  USB  port  for  charging.  If  use  the  other  power  adapter  for 
charging, the output voltage should be 5V DC , the current should be 
greater than 1000mA.
Instrument  displays  charging  animation  when  charging  after  full 
animation  ends,  the  display  is  full  of  symbols.  Charging  time  of  5 
hours.
This instrument adopts lithium ion chargeable battery without memory 
effect and charging can be fulfilled at any time without affecting normal 
operation of the instrument.

 

6

Battery Charging

6

6

6

6

6

6

6

7

Connection method of sensor and drive unit

When installing and remove sensors, turn off the power first.

 

2

2

3

1 sensor   2 Socket of drive unit   3 Drive unit

For  installation,  hold  the  main  body  of  sensor  with  hand,  push  it  into 
socket at the the drive unit as shown in Figure and then slightly pushed 
it to the end of the sheath. To remove, hold the main body of sensor or 
the root of protective sheath with hand and slowly pull it out.

1. Sensor is key part of this tester and great attention.should be paid to 
it.
2.  During  installation  and  remove,  the  stylus  of  sensor  should  not  be 
touched in order to avoid damage and affecting measurement.
3. Connection of sensor should be reliable during installation.
4. The sensor doesn't need to be removed when not in use.
5. Calibration is recommended after each new sensor is installed.

 

8

Connection method of Drive unit and Main Display unit

 

2

Installation method

1  The driver unit is embedded into the main display unit according to 
the direction of arrow 1, so that it is installed on the fixed pin shaft.
2  Press the drive unit in the direction of arrow 1 and down the direction 
of arrow 2 to install the driver unit on the fixed plate.

Summary of Contents for ISR-C300

Page 1: ...MN ISR C300 E www insize com V3 ROUGHNESS TESTER SEPARABLE TYPE OPERATION MANUAL ISR C300 ...

Page 2: ...high capacity no memory effect Can work more than 50 hours while the power is enough Large capacity data storage can store 100 item of raw data and measured profile Real time clock setting and display for easy data recording and storage With automatic sleep automatic shutdown power saving features Reliable circuit and software design of prevent the motor stuck Instrument can display a variety of i...

Page 3: ...lts display 10 Slave results display 11 Battery level 12 Profile scale 13 Profile display area 14 Menu touch key 15 Save touch key 16 Print touch key 5 Buttons define Power key Press and hold START 2 seconds On Off tester Start measurement key Start the instrument to measuring Up Add key Increase the value Cancel Exit key Used to exit the menu and unset Enter key Confirmation setting Down decrease...

Page 4: ... a 2 2 3 1 sensor 2 Socket of drive unit 3 Drive unit For installation hold the main body of sensor with hand push it into socket at the the drive unit as shown in Figure and then slightly pushed it to the end of the sheath To remove hold the main body of sensor or the root of protective sheath with hand and slowly pull it out 1 Sensor is key part of this tester and great attention should be paid ...

Page 5: ...g On Off Press the POWER key to hold 2 seconds after the instrument will automatically boot boot will display type name and manufacturer information and then enter the basic measurement status main display interface In any state press the POWER key to hold 2 seconds after the instrument will be turned off Introductions 1 The next boot will be displayed when the last shutdown set content 2 Startup ...

Page 6: ... measured data to a serial printer The instrument can set the print output according to the actual requirements of arbitrary parameters choose to print or print all the parameters how to set the parameters refer to Print Settings 6 Print measurement results 6 7 Storage measurement results In the main display interface mode press the key to save the measurement results stored in the instrument memo...

Page 7: ...when communicating with mobile APP Bluetooth switch can only be operated when Bluetooth power is off 5 Bluetooth Power Please set the Bluetooth mode first then turn ON the Bluetooth power the instrument will automatically set the Bluetooth module as required Because of the unnecessary battery capacity loss caused by the long term opening of the Bluetooth function the instrument will turn off the B...

Page 8: ... 11 Format memory Data formatting is the deletion of data records Once formatted all data will be emptied Before the data formatting the instrument has the confirmation prompt information After the user confirms the data will not be restored Please proceed with caution Format memory takes about 1 minutes please do not turn off the power 10 Measuring Condition Settings Cut off length λc 0 25mm 0 80...

Page 9: ... the user require high accuracy can be used to correct the indication calibration function and improve measurement accuracy The value of the calibration procedure is as shown Illustration is based on a model calibrated 1 63μm steps to calibrate the model for the actual calibration of the nominal value of the set value 1 Under normal circumstances the instrument in the factory have been rigorously ...

Page 10: ...rned on Connect the instrument to the computer via bluetooth receiver 2 Inset bluetooth receiver receiver to the computer When the green light shows two flashes at the interval the pairing is successful the data can be directly transmitted to Excel by press Enter key 16 Mobile APP This instrument supports the wireless Bluetooth function such as in the high altitude or pipeline operation is inconve...

Page 11: ...eight support feet When the measured surface of the workpiece is smaller than the bases of the instrument the adjustable height support feet can be used as an auxiliary support to complete the measurement 2 Height gauge adapter Options and Usage a Ⅶ Ⅷ Height gauge adapte Protecting Nosepiece Drive unit Height Gauge 3 Extension rod 50mm Extension rod increases the depth for pickup to enter the part...

Page 12: ...sensor operation method The skid of the extra small hole sensor is behind the stylus When it contacts the workpiece the pickup position is first high and then low The use of extra small hole sensors must be installed on the measuring stand for measurement With deep groove sensor it is possible to measure groove with width wider than 3mm and depth deeper than 10mm or the surface roughness of step w...

Page 13: ...olution The Z axis vertical 0 002μm 20μm 0 004μm 40μm 0 008μm 80μm 0 02μm 160μm Measurement item Evaluation Parameter Ra Rz Rq Rt Rc Rp Rv R3z R3y Rz JIS Ry Rs Rsk Rku Rmax Rsm Rmr RPc Rk Rpk Rvk Mr1 Mr2 Standard ISO4287 ANSI b46 1 DIN4768 JISb601 Graphic Primary profile Roughness profile load curves LCD dimensions 3 5 inch 480 320 Filter RC PC RC Gauss D P The sampling length lr 0 25 0 8 2 5mm As...

Page 14: ... Standard Sample Plate The surface of a standard sample plate must be kept clean To avoid scratches on the surface of sample area Troubleshooting When the tester breaks down handle the troubles according to measures described on Fault Information If troubles still exist please return the instrument to factory for repair Users should not dismantle and repair the device by themselves Returned instru...

Page 15: ...viness D P direct profile adopt central line of Least Square Algorithm RC filter analogue 2RC filter with phase difference PC RC filter RC filter with phase correction Gauss filter ISO11562 Traversing Length Parameters Definitions Arithmetical Mean Deviation of Profile Ra Ra is arithmetic mean of the absolute values of profile deviation Z x from mean within sampling length Z1 Z2 Zi Zn Root mean sq...

Page 16: ... the mean line and depth Zv of the deepest profile valley from the mean line within sampling length Total Peak to valley Height Rt Rt is the sum of the height of the highest peak Zp and the depth of the deepest valley Zv over the evaluation length Recommended table of the sampling length 1 1 1 3 ...

Reviews: