Measuring Range
P a ra m e te r
M e a s u rin g ra n g e
R a
R q
0 .0 05 μ m ~ 3 2μ m
R z
R 3 z
R y
R t
R p
R m
0 .0 2μ m ~ 32 0μ m
S k
0 ~ 1 0 0 %
S
S m
1 m m
tp
0 ~ 1 00 %
General Maintenance
probe
Any time swap probes are to be especially careful, careful not to
touch the guide head and a stylus, because this is a key part of the
whole instrument, to try to hold the probe guide head bracket roots
(the front of the body) plug.
To complete the measurement work, please timely probe into the
box;
3. Please pay attention to protect the needle part measuring probe.
The probe's precision components, any knock, touch, fall off
phenomenon may damage the probe, should try to avoid such
situations.
The probe is a damageable parts, do not belong to the scope of
warranty parts, only provide repair. In order not to affect the
measurement work, users are advised to buy backup probe used
for emergency.
1
Main Unit
The instrument is a precision measuring instrument, should
always be handled with care, to avoid the shock.
Pay attention to add oil to maintain regularly every year
to
prevent internal wear.
Battery
Always observe the battery prompt, when the low voltage,
please charge.
The charging time is 3 hours, try not to long time charge.
Standard Sample Plate
The surface of a standard sample plate must be kept clean.
To avoid scratches on the surface of sample area.
Troubleshooting
When the tester breaks down, handle the troubles according to
measures described on Fault Information. If troubles still exist,
please return the instrument to factory for repair. Users should not
dismantle and repair the device by themselves. Returned
instrument should be accompanied with sample plate attached.
Phenomenon of problem should be explained.
2
3
4
5
Error message
Cause
Solutions method
After the boot, the
display interface is
about 1 minutes,
turn off.
The cable plug is not tight.
Re insert socket
The sound of the
motor can not be
heard after turning
on the tester.
Touch screen failure
Touch Parameter loss
touch screen calibration
In the main display
interface, press ESC key for
6 seconds.
Motor error
Motor stuck
Reboot
Out of Range
1. The measured surface
signal
exceeding
the
measurement range
2. Placed away from the
center of the stylus position
Increase Measuring range
Adjust the Stylus position
No test data
After the boot does not
measure.
The actual measurement:
one time
Measurement
Accuracy
Out of Range
Set the parameter error
Calibration data error
Set the parameter
measurement
Calibrate the tester
Pay attention to maintaining the Main Unit surface clean, often
with a soft dry cloth to clean its surface.