insize ISR-C300 Operation Manual Download Page 14

25

26

Measuring Range

P a ra m e te r

 

M e a s u rin g  ra n g e

 

R a 
R q 

0 .0 05 μ m     ~     3 2μ m

 

R z 

R 3 z 

R y 

R t 

R p 

R m 

0 .0 2μ m     ~     32 0μ m

 

S k 

0     ~     1 0 0 %

 

S m 

1 m m 

tp 

0  ~  1 00 %

 

 

 

 

General Maintenance

probe

    Any time swap probes are to be especially careful, careful not to 
touch the guide head and a stylus, because this is a key part of the 
whole instrument, to try to hold the probe guide head bracket roots 
(the front of the body) plug.
   To complete the measurement work, please timely probe into the 
box;
3. Please pay attention to protect the needle part measuring probe.
      The  probe's  precision  components,  any  knock,  touch,  fall  off 
phenomenon  may  damage  the  probe,  should  try  to  avoid  such 
situations.
   The probe is a damageable parts, do not belong to the scope of 
warranty  parts,  only  provide  repair.  In  order  not  to  affect  the 
measurement  work,  users  are  advised  to  buy  backup  probe  used 
for emergency.

1

Main Unit

    The  instrument  is  a  precision  measuring  instrument,  should 
always be handled with care, to avoid the shock.
    Pay  attention  to  add  oil  to  maintain  regularly  every  year

 

to

 

prevent internal wear.

Battery

          Always  observe  the  battery  prompt,  when  the  low  voltage,   
please charge.
     The charging time is 3 hours, try not to long time charge.

Standard Sample Plate

    The surface of a standard sample plate must be kept clean.
    To avoid scratches on the surface of sample area.

Troubleshooting

When  the  tester  breaks  down,  handle  the  troubles  according  to 
measures  described  on  Fault  Information.  If  troubles  still  exist, 
please return the instrument to factory for repair. Users should not 
dismantle  and  repair  the  device  by  themselves.  Returned 
instrument  should  be  accompanied  with    sample  plate  attached. 
Phenomenon of problem should be explained.

2

3

4

5

Error message 

Cause 

Solutions method 

After the boot, the 

display interface is 

about 1 minutes,  

turn off.

 

The cable plug is not tight. 

Re insert socket 

The sound of the 
motor can not be 

heard after turning 

on the tester. 

Touch screen failure 

Touch Parameter loss 

touch screen calibration 

In the main display 

interface, press ESC key for 

6 seconds. 

Motor error 

Motor stuck 

Reboot 

Out of Range 

1.  The  measured  surface 

signal 

exceeding 

the 

measurement range 

2.  Placed  away  from  the 

center of the stylus position 

Increase Measuring range 

 

Adjust the Stylus position 

No test data 

After the boot does not 

measure. 

The actual measurement: 

one time 

Measurement 

Accuracy 

Out of Range 

Set the parameter error 
Calibration data error 

Set the parameter 

measurement 

Calibrate the tester 

 

 

 

 

      Pay  attention  to  maintaining  the  Main  Unit  surface  clean,  often 
with a soft dry cloth to clean its surface.

Summary of Contents for ISR-C300

Page 1: ...MN ISR C300 E www insize com V3 ROUGHNESS TESTER SEPARABLE TYPE OPERATION MANUAL ISR C300 ...

Page 2: ...high capacity no memory effect Can work more than 50 hours while the power is enough Large capacity data storage can store 100 item of raw data and measured profile Real time clock setting and display for easy data recording and storage With automatic sleep automatic shutdown power saving features Reliable circuit and software design of prevent the motor stuck Instrument can display a variety of i...

Page 3: ...lts display 10 Slave results display 11 Battery level 12 Profile scale 13 Profile display area 14 Menu touch key 15 Save touch key 16 Print touch key 5 Buttons define Power key Press and hold START 2 seconds On Off tester Start measurement key Start the instrument to measuring Up Add key Increase the value Cancel Exit key Used to exit the menu and unset Enter key Confirmation setting Down decrease...

Page 4: ... a 2 2 3 1 sensor 2 Socket of drive unit 3 Drive unit For installation hold the main body of sensor with hand push it into socket at the the drive unit as shown in Figure and then slightly pushed it to the end of the sheath To remove hold the main body of sensor or the root of protective sheath with hand and slowly pull it out 1 Sensor is key part of this tester and great attention should be paid ...

Page 5: ...g On Off Press the POWER key to hold 2 seconds after the instrument will automatically boot boot will display type name and manufacturer information and then enter the basic measurement status main display interface In any state press the POWER key to hold 2 seconds after the instrument will be turned off Introductions 1 The next boot will be displayed when the last shutdown set content 2 Startup ...

Page 6: ... measured data to a serial printer The instrument can set the print output according to the actual requirements of arbitrary parameters choose to print or print all the parameters how to set the parameters refer to Print Settings 6 Print measurement results 6 7 Storage measurement results In the main display interface mode press the key to save the measurement results stored in the instrument memo...

Page 7: ...when communicating with mobile APP Bluetooth switch can only be operated when Bluetooth power is off 5 Bluetooth Power Please set the Bluetooth mode first then turn ON the Bluetooth power the instrument will automatically set the Bluetooth module as required Because of the unnecessary battery capacity loss caused by the long term opening of the Bluetooth function the instrument will turn off the B...

Page 8: ... 11 Format memory Data formatting is the deletion of data records Once formatted all data will be emptied Before the data formatting the instrument has the confirmation prompt information After the user confirms the data will not be restored Please proceed with caution Format memory takes about 1 minutes please do not turn off the power 10 Measuring Condition Settings Cut off length λc 0 25mm 0 80...

Page 9: ... the user require high accuracy can be used to correct the indication calibration function and improve measurement accuracy The value of the calibration procedure is as shown Illustration is based on a model calibrated 1 63μm steps to calibrate the model for the actual calibration of the nominal value of the set value 1 Under normal circumstances the instrument in the factory have been rigorously ...

Page 10: ...rned on Connect the instrument to the computer via bluetooth receiver 2 Inset bluetooth receiver receiver to the computer When the green light shows two flashes at the interval the pairing is successful the data can be directly transmitted to Excel by press Enter key 16 Mobile APP This instrument supports the wireless Bluetooth function such as in the high altitude or pipeline operation is inconve...

Page 11: ...eight support feet When the measured surface of the workpiece is smaller than the bases of the instrument the adjustable height support feet can be used as an auxiliary support to complete the measurement 2 Height gauge adapter Options and Usage a Ⅶ Ⅷ Height gauge adapte Protecting Nosepiece Drive unit Height Gauge 3 Extension rod 50mm Extension rod increases the depth for pickup to enter the part...

Page 12: ...sensor operation method The skid of the extra small hole sensor is behind the stylus When it contacts the workpiece the pickup position is first high and then low The use of extra small hole sensors must be installed on the measuring stand for measurement With deep groove sensor it is possible to measure groove with width wider than 3mm and depth deeper than 10mm or the surface roughness of step w...

Page 13: ...olution The Z axis vertical 0 002μm 20μm 0 004μm 40μm 0 008μm 80μm 0 02μm 160μm Measurement item Evaluation Parameter Ra Rz Rq Rt Rc Rp Rv R3z R3y Rz JIS Ry Rs Rsk Rku Rmax Rsm Rmr RPc Rk Rpk Rvk Mr1 Mr2 Standard ISO4287 ANSI b46 1 DIN4768 JISb601 Graphic Primary profile Roughness profile load curves LCD dimensions 3 5 inch 480 320 Filter RC PC RC Gauss D P The sampling length lr 0 25 0 8 2 5mm As...

Page 14: ... Standard Sample Plate The surface of a standard sample plate must be kept clean To avoid scratches on the surface of sample area Troubleshooting When the tester breaks down handle the troubles according to measures described on Fault Information If troubles still exist please return the instrument to factory for repair Users should not dismantle and repair the device by themselves Returned instru...

Page 15: ...viness D P direct profile adopt central line of Least Square Algorithm RC filter analogue 2RC filter with phase difference PC RC filter RC filter with phase correction Gauss filter ISO11562 Traversing Length Parameters Definitions Arithmetical Mean Deviation of Profile Ra Ra is arithmetic mean of the absolute values of profile deviation Z x from mean within sampling length Z1 Z2 Zi Zn Root mean sq...

Page 16: ... the mean line and depth Zv of the deepest profile valley from the mean line within sampling length Total Peak to valley Height Rt Rt is the sum of the height of the highest peak Zp and the depth of the deepest valley Zv over the evaluation length Recommended table of the sampling length 1 1 1 3 ...

Reviews: