background image

3

 Specifications

Measure Range: 0 100HD

~

Measure accuracy: 

1HD

±

Data capacity: 500
Voltage: 3.6V (rechargeable Ni-MH battery)
Continuous working time by one time charge: about 20 hours
Battery charge time: about 3 hours
Work temperature: 0°C~+50°C
Work humidity: 20%~85%
Storage temperature: -30°C~+80°C
Storage humidity: 5%~95%
Outline size: 153mm×50mm×29mm (main body)
Weight: about 171g
Strut diameter of the needle: (1.25±0.10) mm
The top plane diameter of the needle: (0.10±0.01) mm
The top cone angle of the needle: Initial verification: (30±0.5)°

Operation Instructions

1.  Button and measure interface

 1) Button

Power/Return: By pressing this button, the durometer will be power

 on. By long time pressing this button, the durometer will be power  
 off. By pressing this button, the screen will return to the upper level 

 menu.

 pressing this button. Under the menu interface, it will enter the next 
 menu interface. In addition, it is the confirmation button under function
 selection or parameters setting interface.

 OK: Under the measure interface, it will enter the menu interface by 

Fig 3-1

14

 

 

Low-voltage alarming and recharge

 The durometer will be recharged by the packaged charger. The  

durometer will not be auto power on during recharging. The charging  
status will show on the measuring interface by hand power on. The  
moving of the battery level icon line indicate it is charging now. If the  
icon fill by all line, it means the charging process is finished. If the power 
is low, the battery level icon will flash to prompt to be recharged in time.

 Reset

Software reset: if the software abnormal is happened, the durometer  
will reset automatically.
Hardware reset: the durometer can be reset by the hardware reset  
under the crash. The reset button is local in the hole of the back shell.

Data printing

Digital durometer can connected to the special mini-thermal printer to 
print the hardness result report. The mini-thermal printer is optional part. 
If it is needed, please contact the distributor.
The printer should be put near to the durometer and the distance  
shouldn't be over 3m.

1. Format of the report

     A full report is as the figure 7-1. Ps:1.the print content can be 

customized by the relative menu explain on above.

          The date/time in the report content is the measuring date/time. The 

date/time at the end of the report is the print date/time.

Fig 7-1

Summary of Contents for ISH-DSD

Page 1: ...ISH DSD www insize com OPERATION MANUAL DIGITAL SHORE DUROMETER MN ISH DSD E V0...

Page 2: ......

Page 3: ...foot surface The value of L indicates the value of Shore Hardness of the sample Larger the value of L smaller the Shore Hardness Conversely smaller the value of the L sample s Larger the sample s Shor...

Page 4: ...n flooring bowing Especially for on site hardness measurement of rubber and plastic products 15 2 Print report There are several print modes built in the durometer You can select by yourself 3 Auto pr...

Page 5: ...Low voltage alarming and recharge The durometer will be recharged by the packaged charger The durometer will not be auto power on during recharging The charging status will show on the measuring inter...

Page 6: ...e current times Measure mode Indicating the current measure mode AVE Average measure mode Max Maximum value measure mode MAX AVE Maximum average value measure mode 2 Menu 7 6 5 4 3 2 1 Fig 4 1 Hardnes...

Page 7: ...r this menu All data will be deleted by selection of YES and this operation will be aborted by selection of NO Fig 4 22 5 Print setting The data saved in the durometer can be printed by the thermal pr...

Page 8: ...ssing button UP The measure date time sample code of the current data can be shown pressing button OK 6 1 Calibration Pressing button OK to enter the menu select Calibration press button The durometer...

Page 9: ...tically The average value of maximum value of every measuring will be calculated and this value will be taken as the last result Measure times setting The statistic times can be set under this menu Th...

Page 10: ...l be button RETURN be aborted by press 8 Tolerance limit setting Under this menu the tolerance limit can be set When the measuring data is exceed the setting value the icon of will be showed on the LC...