insize ISH-DSD Operation Manual Download Page 8

11

 ④ 

Alarm sound setting

 If this function is set as on, the durometer will make a sound of 
 alarm when the data is exceed the tolerance limit.
 The selection will be modified by pressing button “UP” or “DOWN”. 

The selection will be confirmed by pressing button 

. The 

“OK”

selection will be aborted by press button “RETURN”.

Fig 4-18

 

 

 

 

4) Storage management

① 

Browse from the first data

 The first page data will be show when enter this menu (8 data per 
 page). It will turn to the next page by pressing button “DOWN”.  
 The selection will be shifted by pressing button “UP”. The measure 
 date/time/sample code of the selected data can be show by
 pressing button “OK”.

Fig 4-19

Fig 4-20

 

 

Browse from the last data

The last page data will be show when enter this menu (8 data per 
page). It will turn to the previous page by pressing button “UP”. The 
selection will be shifted by pressing button “DOWN”. The measure 
date/time/sample code of the current data can be show by pressing 
button “OK”.

 

 

  

 Browse the selected data

The range of data to be browse should be set up under this menu. 
The first page of selected data will be show by pressing button 
“OK”. It will turn to the next page by pressing button “DOWN”. The 
selection will be shifted by pressing button “UP”. The measure 
date/time/sample code of the current data can be shown pressing
button “OK”.

6

 1) Calibration

 Pressing button “OK” to enter the menu, select “Calibration”, press 
 button 

 The durometer can be calibrated by glass plate,

“OK”.

 standardized thickness block or other rubber whose hardness is 
 known. Make the press needle of the durometer press on the glass 
 plate, standardized thickness block or other rubber whose hardness 
 is known. Keeping the presser foot contact the samples closely for a 
 while to stabilize, the LCD show the hardness value. If the hardness 
 value is different from the theory value (calibration value), please 
 enter the theory value. If the glass plate is applied, please enter 100. 
 If the standardized thickness block is applied, please enter 50. If 
 the rubber whose hardness is known is applied, please enter the 
 rubber's hardness.
 Values enter method: under the enter value status, the corresponding 
 position will flash, press ”UP” to adjust the number, 

 

press “DOWN” to

 shift to the next position, press  RETURN” to abort 

  

the calibration,

 press “OK” to confirm the calibration.

 2) Measure setting

 Under measure setting, the measure mode, measure times, tolerance
 limit, on or off of gross error processing setting, sample code can be
 set. The first page of measure setting menu is showed in 

 

figure 4-6.

 The second page of measure setting menu is showed in 

 

figure 4-7.

 

Fig 4-5

 

Fig 4-6

Fig 4-7

Summary of Contents for ISH-DSD

Page 1: ...ISH DSD www insize com OPERATION MANUAL DIGITAL SHORE DUROMETER MN ISH DSD E V0...

Page 2: ......

Page 3: ...foot surface The value of L indicates the value of Shore Hardness of the sample Larger the value of L smaller the Shore Hardness Conversely smaller the value of the L sample s Larger the sample s Shor...

Page 4: ...n flooring bowing Especially for on site hardness measurement of rubber and plastic products 15 2 Print report There are several print modes built in the durometer You can select by yourself 3 Auto pr...

Page 5: ...Low voltage alarming and recharge The durometer will be recharged by the packaged charger The durometer will not be auto power on during recharging The charging status will show on the measuring inter...

Page 6: ...e current times Measure mode Indicating the current measure mode AVE Average measure mode Max Maximum value measure mode MAX AVE Maximum average value measure mode 2 Menu 7 6 5 4 3 2 1 Fig 4 1 Hardnes...

Page 7: ...r this menu All data will be deleted by selection of YES and this operation will be aborted by selection of NO Fig 4 22 5 Print setting The data saved in the durometer can be printed by the thermal pr...

Page 8: ...ssing button UP The measure date time sample code of the current data can be shown pressing button OK 6 1 Calibration Pressing button OK to enter the menu select Calibration press button The durometer...

Page 9: ...tically The average value of maximum value of every measuring will be calculated and this value will be taken as the last result Measure times setting The statistic times can be set under this menu Th...

Page 10: ...l be button RETURN be aborted by press 8 Tolerance limit setting Under this menu the tolerance limit can be set When the measuring data is exceed the setting value the icon of will be showed on the LC...

Reviews: