10. Technical data
(according to standards IEC 60601-1, IEC 60601-2-7)
Modell
350HF
450HF
550HF
650HF
850HF
@ 80kV
300 mA
400 mA
500 mA
600 mA
800 mA
@ 100kV
300 mA
320 mA
400 mA
500 mA
630 mA
Maximum anode current
in radiological mode
@ 150kV
200 mA
200 mA
250 mA
250 mA
320 mA
Maximum anode current in fluoroscopic
mode (40-125kV)
10mA
Maximum anode current in radiological
mode
(X-Ray tube voltage)
300 mA
(100kV)
400 mA
(80kV)
500 mA
(80kV)
600 mA
(83kV)
800 mA
(81kV)
X-Ray parameters resulting the maximum
output power in radiological mode
100kV,
300mA
80kV,
400mA
80kV,
500mA
100kV,
500mA
103kV,
630mA
X-Ray parameters resulting the maximum
output power in fluoroscopic mode
125kV, 10mA
Maximum output power
(parameter combinations)
30 kW
(100kV,
300mA,
100ms)
32 kW
(80kV,
400mA,
100ms)
40 kW
(80kV,
500mA,
100ms)
50 kW
(100kV,
500mA,
100ms)
65 kW
(103 kV,
630mA,
100ms)
Minimum current time product
0.5mAs
Nominal shortest irradiation time in
AEC mode
5ms