TOP-X100NR USER MANUAL
Innomed Medical Inc.
35
Three different image storage operating modes can be set: single, “burst” and series.
In
burst operating mode
the generator makes 10 images with the set frequency, than
returns to fluoroscopy mode. In
series operating mode
the images are made with the
set frequency as long as the REC button is pressed. Obviously one image is made in
single operating mode
. The image creation frequency is increased and decreased
with buttons 8 and 9. The adjustable image frequencies are set at the installation
time.
8.5 Fluoroscopic parameter setting on the control panel
All the fluoroscopic parameter setting, which are possible on the remote controller
are also available on the console. Further possibilities are described in chapter 8.4.
8.5.1 Displaying fluoroscopic parameters on the control panel
With the exception of the APR screen, the fluoroscopic parameters are always visible
on the control panel.
On the left side of the display is the x-ray tube fluoro voltage; on the right, the
fluoroscopic x-ray current; and in the center, the other fluoro parameters are visible.
The following parameters appear in the middle box:
Image content inverting
Image horizontal mirroring
Image vertical mirroring
If these symbols appear on the display, they indicate active functions; if they do not
appear, the appropriate functions are disabled.
To the right of these indications is the digital image storage fluoroscopic operating
mode:
Single Spot – Digital storage of a single image