TOP-X100NR USER MANUAL
Innomed Medical Inc.
22
Time steps [ms or s]:
1, 2, 3, 4, 5, 6, 8, 10, 12, 16, 20, 25, 32, 40, 50, 63, 80, 100,
120, 160, 200, 250, 320, 400, 500, 630, 800 ms,
1, 1.2, 1.6, 2, 2.5, 3.2, 4, 5, 6.3, 8, 10 s
7.1.3 mAs mode (or two-point mode)
Using the mAs up (37) / mAs down (36) buttons changes the mAs value. The kV
value will not change, but the mA and time parameters will. According to exposure
parameters the generator will try to keep the exposure time around the value set at
installation (typically 100ms), if load and current limits make it possible. When
pushed and held down, the mAs value increases according to the following steps:
mAs steps [mAs]:
0.5, 0.6, 0.8, 1, 1.2, 1.6, 2, 2.5, 3.2, 4, 5, 6.3, 8, 10, 12, 16, 20, 25, 32,
40, 50, 63, 80, 100, 125, 160, 200, 250, 320, 400, 500, 630, 800, 1000
7.1.4 PATIENT – Patient size
Patient size is mainly used to correct parameters set in APR mode.
If the patient is fatter or thinner than average, or the patient is a child, use one of the
patient buttons (21-24). In this case you can see that the exposure voltage (kV)
changes; when the child button is used, mAs also changes (increases in case of
fatter patient, decreases in case of child or thinner patient). The other parameters
remain unchanged. The buttons have the same functions in free parameter setting
operating mode as well.
If the required change cannot be set due to load limits, the LED indicating patient
size does not turn on, and the previous selection remain.