TOP-X100NR USER MANUAL
Innomed Medical Inc.
15
5. Anatomical program mode
After turning on, the generator starts with the
APR
operating mode – a low-power
program.
In the APR operating mode of the
TOP-X 100NR
12 buttons (9-20) are reserved in
the APR field (A) for the anatomical programs. Each button is an arbitrary level,
independent anatomical program. During installation it is possible to set each field’s
unique name.
When a body part button (9-20) is pushed, the selected body part’s sub-menu is
displayed. Here 12 body parts or sub-body parts selections are available.
When the anatomical program is displayed (sub body part selection including
settings), all program-related settings appear on the control panel displays.
Selecting an anatomical program (APR) always means the automatic setting of the
next parameters (in this example at the “SKULL AP”).
kV:
70kV
mAs:
40mAs
Device (TECHNIQUE):
Vertical Bucky stand
Grid use (GRID):
Grid is used
High-speed starter (HS):
Not used
Tube (TUBE):
Tube 1
Focal spot (FOCUS):
Large focal spot
AEC ON/OFF:
On
Dominant field (FIELD):
2. (middle) is selected
Film type (SCREEN):
M medium film sensitivity
Density correction (DENSITY):
"0" (middle setting)
Patient thickness (PATIENT):
Normal (average) size
Load (LOAD 100%):
Allows max. 80% load
The parameter values retrieved in APR mode are preset to a patient with average
height and weight, but the parameters can be changed anytime.