In impulse operating mode the generator enables the radiation in synch with the
square signal released by the camera. Even lower dosage is possible by emitting
less than 25 radiation impulse per second, but this will decrease the continuity of
motion. To modify the impulse frequency press and hold the impulse button (9) in
active impulse mode. The frequency will increase or decrease, release the button
when the desired frequency is reached. If you press and hold the impulse button (9)
again, the frequency will change in the other direction. The adjustable impulse
frequency steps are set at installation.
8.4.7 Automatic brightness setting (ABS)
Based on the brightness signal received from the TV chain and independently from
the radiation absorption of the examined organ, the system can display stable image
brightness on the monitor. After you have calibrated the controller with the required
brightness level, push the ABS button (5) to activate this function.