DOCUMENT NO. 920012 REV 2 EFFECTIVE 03-14-18 INFRASCAN CONFIDENTIAL PAGE 13 OF 35
The Disposable Fiber Optic Shield is used to couple the light to the
patient’s head.
Attaching the Shield turns the Infrascanner on, and removing the Shield turns the
Infrascanner off.
Caution: Select a Shield for each patient. Do not re-use the Disposable Fiber Optic
Shield on a different patient. The Disposable Shield is for single patient use only.
When the Disposable Shield contacts the patient, it is capable of transferring
infectious agents.
Note: The Shield is non-sterile and should not be sterilized.
Note: When reusing on same patient, the Shield can be cleaned by wiping with a soft
tissue damp with alcohol.
Note: Dispose of used Shields in accordance with local ordinances.
3.0
OPERATING PROCEDURE
3.1
SETTING UP THE SYSTEM
3.1.1.
Battery Installation. Use a ¼” flat blade screwdriver (or a coin) to release the
battery door latch by turning one half turn counter clockwise.
LED’s and Cradle
ON/OFF switch
Figure 2-6: Model 2000 Cradle
Figure 2-7: Model 2000 Disposable Shield