DOCUMENT NO. 920012 REV 2 EFFECTIVE 03-14-18 INFRASCAN CONFIDENTIAL PAGE 15 OF 35
Infrascanner turns on when the Infrascanner is placed in the Cradle, disconnect the
mains power briefly and reconnect.
3.1.3.1 If the disposable batteries are installed, the amber Charge LED will not
illuminate continuously. If the rechargeable battery pack is installed, after the
variable time of blinking, the amber
“Charge” LED will illuminate continuously.
If no batteries are installed after the 10 seconds of blinking, neither the Charge
LED nor the Fault LED will illuminate.
3.1.3.2 The Infrascanner requires approximately 6 hours to charge, and is fully
charged when the amber
“Charge” LED on the Cradle extinguishes. The
Infrascanner battery will not charge if the Infrascanner is ON or if the battery
temperature is below 0 degrees Celsius or above 50 degrees Celsius. The
Infrascanner detects when disposable batteries are in the Infrascanner and will
not charge them. There is no hazard under this condition. If the orange Fault
LED illuminates replace the rechargeable battery pack.
3.1.4.
Fiber Optic Disposable Shield. Press fibers outward so that don’t get bent
against the windows during installation, as shown in Figure 3-3.
3.1.5.
Attach the Disposable Shield to the Infrascanner as shown in Fig. 3-4. Place
the fibers over the optical towers and push firmly at the top/back of the disposable,
until you feel the disposable snap onto the Infrascanner. Once the disposable
Shield is installed press the fibers inward so that they are up against the windows.
Note: Attachment of the Fiber Optic Disposable Shield turns on the
Infrascanner.
Caution: Select a new Disposable for each patient. Do not reuse the
Disposable Fiber Optic Shields on a different patient. Disposable Shields
contact the patient and may be contaminated. The Disposable Shield is
Single Patient Use.
Figure 3-3: Proper Placement of Fibers