User’s Manual for Omega 12Patient Monitor
165
C Alarm Messages
C.1 Physiological Alarm Messages
General Physiological Alarm
Alarm Messages
Default
Level
Cause and Solution
XXX Too High
Medium
XX value has risen above the high alarm limit
or fallen below the low alarm limit. Check the
patient’s condition and check if the patient
category and alarm limit settings are correct.
XXX Too Low
Medium
Resp Apnea
High
Patient can’t respirate or the respiration signal
was so weak that the monitor cannot perform
respiration analysis. Check the patient’s
Condition, electrode placement and the Resp
connections.
Note: XXX represents Physiological parameter, like HR, ST, RR, SpO2, PR, EtCO2,
ART-SYS, ART-DIA etc.
Arrhythmia Alarm messages
Alarm Messages
Default Level
Asystole
High
V-Fib/V-Tach
Medium
V-Tach
Medium
R on T
Medium
RUN
Medium
Vent Rhythm
Medium
Couplet
Medium
PVC
Medium
Bigeminy
Medium
Trigeminy
Medium
Tachy
Medium
Brady
Medium
Missed Beats
Medium
C.2 Technical Alarm Messages
Module
Alarm Messages
Default
Level
Cause and Solution
ECG
ECG Lead Off
Low
The electrode has become