User’s Manual for Omega 12Patient Monitor
156
Co2 Resolution
在
0 - 69mmHg: 0.1mmHg
在
70 - 150mmHg: 0.25mmHg
Accuracy*
0 - 40mmHg: ±2mmHg
41 - 70mmHg: ±5%
71 - 100mmHg: ±8%
101 - 150mmHg: ±10%
respiration rate> 80/min: ±12%
*NOTE: the above data is based on gas temperature of 25℃.
Accuracy Drift
Short-term drift:< 0.8mmHg in 4-hour operation
Long-term drift: Meet the requirement for measurement
accuracy within 120 hours
Noise
At 5% CO2 concentration,the Root Mean Square (RMS) of
sensor noise is not greater than 0.25 mmHg.
Frequency
100Hz
awRR Measurement Range
2 - 150/min (rpm)
awRR Measurement
Precision
±1rpm
Alarm Limit
Range
EtCO2 Alarm High
(Low limit + 2) ~150mmHg manual input required
Default value: 60mmHg
EtCO2 Alarm Low
0~ (high limit - 2) mmHg manual input required
Default value: 15mmHg
awRR Alarm High
(Low limit + 2) ~150rpm manual input required
Default value:
30rpm (Adult)
100rpm (Neonate)
awRR Alarm Low
0~ (high limit - 2) rpm manual input required
Default value:
5rpm (Adult)
30rpm (Neonate)
A.5.9 BISx2