User’s Manual for Omega 12Patient Monitor
129
21.8 Loading Paper
1.
Press the door latch to open the recorder door. Remove the paper core. Place a new
paper roll between the tabs of the paper holder.
2.
Make sure the paper unrolls from underneath the paper roll. Close the recorder door,
make sure the new paper roll fits snugly into its housing and the paper end is feeding
from the top. Otherwise, it would be hard to feed the paper under the roller.
21.9 Removing Paper Jam
If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam
first. If a paper jam is detected, follow this procedure to remove it:
1.
Open the recorder door.
2.
Take out the paper and tear off the draped part.
3.
Reload the paper and close the recorder door.
NOTE
Use only specified thermal paper. Otherwise, it may cause damage to the recorder’s printhead,
the recorder may be unable to print, or poor print quality may result.
Never pull the recorder paper with force when a recording is in process. Otherwise, it may
cause damage to the recorder.
Do not leave the recorder door open unless you reload paper or remove troubles.