How to Set your IRwin Methane Leak Detector
for 1 Point Calibration 2.5%
7
Use to select
Generic
and press
followed by
8
9
Use and to navigate to
Calibration
and press
Use
to navigate to
Settings
and press
10
If prompted, enter the (min) Intermediate password code
If not prompted, go ahead to next step
6
If prompted, enter the Supervisor (Advanced) password code
If not prompted, go ahead to next step
10