No.
Message
Possible error sources
Remedy
E5
Battery level too low
(count down to power
off)
Battery low
Switch off detector and charge battery
E7
SnO
2
voltage error
System error
Restart detector.
If problems remains: Contact INFICON service.
E8
Detector shuts down
System error
Restart detector.
If problems remains: Contact INFICON service.
E9
Red LED and constant
beep
HW watchdog trigged
Perform hardware reset.
Restart detector.
If problems remains: Contact INFICON service.
E10
O
2
Sensor bias lost
O2 sensor bias lost due
to uncontrolled shut
down.
Check in Ex Tox mode that O
2
value is OK (close
to 20.9%). If not, leave the detector on until O2
sensor shows 20 - 22% in air. This can take up to
12 hours.
E11
“IR Lamp" has red
cross in function test
result screen.
IR system error
Restart detector.
If problems remains: Contact INFICON service.
E12
“Pump & Valve" test
has red cross in
function test result
screen.
Exhaust blocked
Restart detector.
If problems remains: Contact INFICON service.
Pump dirty or broken
Contact INFICON service.
E13
“Pump & Valve" test
has red cross in
function test result
screen.
Valve clogged
Restart detector.
If problems remains: Contact INFICON service.
System error
Restart detector.
If problems remains: Contact INFICON service.
E14
“Pump & Valve" test
has red cross in
function test result
screen.
Exhaust blocked
Check that nothing is blocking the exhaust.
If alarm remains: Contact INFICON service.
Reference inlet blocked
Check that nothing is blocking the reference inlet.
If alarm remains: Contact INFICON service.
Pump dirty or broken
Contact INFICON service.
E15
"Gas Sensors" has red
cross in function test
result screen.
Sensor error or warm-up
drift
Restart detector.
If problems remains: Contact INFICON service.
E16
"Gas Sensors" has red
cross in function test
result screen.
Sensor error or warm-up
drift
Restart detector.
If problems remains: Contact INFICON service.