3 - 33
IP
N 07
4-
39
7-
P1
G
HAPSITE Smart Operating Manual
3.6 GC/MS Mode with Tenax Concentrator
This method is used for analyzing samples with concentration levels in the low part
per billion to high part per trillion range. This method is used when the loop method
fails to detect a chemical and is similar to the Tri-Bed concentrator method. The
Tenax concentrator will not effectively concentrate compounds with boiling points
below 80 degrees Centigrade. Refer to
, for additional
information on GC/MS methods.
NOTE:
Before a GC/MS can be run in this Concentrator mode, the Tenax
concentrator must be installed. Refer to
Section 2.9.4, Installing the Tenax
or Carbopack Concentrator, on page 2-31
for instructions. Once installed,
the Tenax concentrator must be cleaned before sampling begins.
1
From the main menu on the Front Panel display of the HAPSITE select
Run
Method
by highlighting the selection and pressing
SEL
.
Figure 3-44 Main Menu on Front Panel
2
Use the up or down arrows to highlight the
GCMS
option on the
Method
menu.
Press
SEL
. See
.
Figure 3-45 Selecting GCMS using the Method Menu
Summary of Contents for HAPSITE Smart
Page 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Page 2: ......
Page 12: ......
Page 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Page 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Page 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 548: ......