HAPSITE Smart Operating Manual
5.6.1 Quick Reference SOP — GC/MS Mode with HSS and Sample Loop
1
Double-click the
Smart IQ
software icon.
2
Double-click the
Run Method
icon.
3
Double-click the
GCMS
folder.
4
Double-click the
Headspace
folder.
5
Double-click the
HSS 15 Min Loop
method.
6
Wait for heaters to gain control of set temperatures
7
Following a 20 minute sample temperature equilibration in the HSS oven,
gently depress the headspace needle into the 40 mL sample vial.
8
Click the
Run
button in the pop-up window.
9
When prompted, place the headspace needle in the purge vial and press
Sel
.
10
To analyze the data, see
5.7 GC/MS Mode with Headspace Sampling System and
Concentrator
This method is used for analyzing samples with concentration levels in the low part
per billion to high part per trillion range. This method is used when the Headspace
Sampling System with the Sample Loop method fails to detect a chemical. Refer
to
for additional information on GC/MS methods.
CAUTION
The concentrator feature is designed to enhance
HAPSITE sensitivity when analyzing low concentration
samples. The accumulated sample VOCs are transferred
from the concentrator adsorbent bed to the HAPSITE as
a concentrated sample. Care should be taken to avoid
overloading the HAPSITE and causing excessive
carryover of VOCs in subsequent runs.
1
Refer to
Chapter 13, Headspace Sampling System
for more information on the
Headspace System. Refer to
Section 2.6, Headspace Sampling System, on
for assembly instructions.
2
Follow the instructions in
Section 2.9.3, Installing the Tri-Bed Concentrator, on
Section 2.9.4, Installing the Tenax or Carbopack Concentrator, on
to install the concentrator tube.
Summary of Contents for HAPSITE Smart
Page 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Page 2: ......
Page 12: ......
Page 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Page 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Page 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 548: ......