15 - 39
IP
N 07
4-
39
7-
P1
G
HAPSITE Smart Operating Manual
Figure 15-43 Service Window
4
Select
Activate Neg
.
5
When prompted on the Front Panel, enter the
Serial Number
of the New NEG
Pump.
6
The NEG Pump will take approximately 18 hours to activate. The process is
totally automated and does not need any other interactions until the process is
complete.
7
Once the NEG Pump is activated, remove the HAPSITE from the Service
Module. (Power the HAPSITE with battery or an external power supply.)
Refer to
section 14.8, Detaching the HAPSITE, on page 14-17
for instructions
on safely removing the HAPSITE from the Service Module.
NOTE:
When completed, the Activate NEG process will automatically detach the
HAPSITE from the Service Module.
8
Repeat the leak check portion of this process. Instead of checking for a leak
between the Analytical Module and Service Module, check the
VI
valve on
the bottom of the HAPSITE where the Service Module physically attaches.
(Refer to
section 15.7.5, Part 4: Leak Check of the Vacuum System, on page
.)
9
If there is a leak, the leak must be sealed by repeating
1: NEG Pump Removal, on page 15-17
through
section 15.7.4, Part 3: Install
the New NEG Pump, on page 15-27
, the leak check in
Leak Check of the Vacuum System, on page 15-32
repeated and the NEG
Pump reactivated by performing
section 15.7.6, Part 5: Activation of the NEG
.
Summary of Contents for HAPSITE Smart
Page 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Page 2: ......
Page 12: ......
Page 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Page 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Page 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Page 548: ......