tina88e1 (2014-01)
33
Evacuate the gauge to a pressure according to the table
below:
F.S.
Recommended final pressure for
zero adjustment
1000 Torr/mbar
100 Torr/mbar
10 Torr/mbar
1
Torr/mbar
0.1 Torr/mbar
<5×10
-2
Torr
<5×10
-3
Torr
<5×10
-4
Torr
<5×10
-5
Torr
<5×10
-6
Torr
<6.65×10
0
Pa
<6.65×10
-1
Pa
<6.65×10
-2
Pa
<6.65×10
-3
Pa
<6.65×10
-4
Pa
<5×10
-2
mbar
<5×10
-3
mbar
<5×10
-4
mbar
<5×10
-5
mbar
<5×10
-6
mbar
If the final pressure is too high for zero adjustment (>25%
of the F.S.), the zero cannot be reached and the
<STATUS> LED blinks green. If this is the case, activate
the factory setting and adjust the zero again (
→
39).
Operate the gauge for at least 90 minutes under constant
ambient conditions (until the signal is stable).
Briefly press the <ZERO> button with a pin. The zero
adjustment runs automatically. The <ERROR> LED blinks
red until the adjustment (duration
≤
8 s) is completed.
Press the button briefly
max. ø1.1 mm
After zero adjustment, the gauge automatically returns
to the measurement mode.