36
tina88e1 (2014-01)
4.5 Switching
Functions
The two switching functions can be set to any pressure within
the measurement range of the gauge (
→
16).
The current setpoint setting
•
is output at the analog port <TEMP / ATM> instead of the
measurement signal (
→
25) and can be measured with a
voltmeter after the <SETPOINT> button is pressed, or
•
can be read/written via the RS232, Ethernet or wireless
interface.
•
can be read/written via the diagnostic port,
If the pressure is lower than the setpoint, the corresponding LED
(<SETPOINT 1> or <SETPOINT 2>) is lit solid and the corre-
sponding relay (
→
25) is energized.
Measurement signal (pressure p)
Time t
Me
asu
reme
nt value
Off
On
Hysteresis
(1% F.S.)
Off
Threshold value
Setpoint