![Inficon Cube CDGsci Operating Manual Download Page 31](http://html.mh-extra.com/html/inficon/cube-cdgsci/cube-cdgsci_operating-manual_2054966031.webp)
tina88e1 (2014-01)
31
An existing wireless network can be displayed by using
•
the RS232 ASCII command
IPW
, or
•
the REST services command
http://192.168.0.248/1/cmd/IPW
The output is the current IP address and the current network
mask (e.g. 10.0.1.6|255.255.255.0).
4.3 Websites
The gauge is equipped with a web server. Via its websites user
can
•
poll the status of the gauge,
•
query gauge errors,
•
set setpoints, or
•
change communication parameters (e.g. IP address)
URL of the web sites:
http://[IP address]:8087
(e.g.
192.168.0.248:8087)
Internet browser, e.g.
•
Mozilla Firefox, version >20, or
•
Windows Internet Explorer, version >10.0
IP addresses (default)
→
11.
Changing IP addresses
→
28.