tina88e1 (2014-01)
37
4.5.1 Adjusting the Setpoints
The setpoints can be adjusted via
•
the buttons on the gauge,
•
the diagnostic port (
→
[3]),
•
the RS232 interface (
→
[2], [3]),
•
the Ethernet interface (
→
[1]),
•
the wireless network (
→
[4], [5]).
DANGER
DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SETPOINT>
button the measurement signal is suppressed and
the corresponding threshold value is output in-
stead. This can cause malfunctions.
Push the <SETPOINT> button only if you are sure
that no damages can arise from a malfunction.
Adjusting <SETPOINT 1>
Push the <SETPOINT> button with a pin (max. ø1.1 mm).
The gauge changes to the switching function mode and
outputs the current threshold value at the measurement
value output for about 10 s (LED <SETPOINT 1> blinks).
Press the button briefly
max. ø1.1 mm