Operation
[email protected] | romanoff.com
7–10
7.5
VC480V Front panel below display
Figure 18: Front panel below display
Manual control:
Generator
Start
By this button you can start heating from generator. Switched on genera-
tor regulate for the set temperature (see above ‘temperature’).
Generator
Stop
Stops heating (if not operating in program).
Vacuum
Chamber
Opens the vacuum chamber. Chamber will be closed automatically when
it is swivelled completely under the crucible chamber. During granulating
process lifting and lowering of the granulating tank.
Flask
Lift
Lift and sink of flask lift. This function is only possible with closed vacuum
chamber. With open chamber lifting and sinking is performed automati-
cally.
Sealing
Rod
With pressing this button the sealing rod opens or close, which let flow
liquid material in the flask or granulation tank when it is requested. With
software 800.0104 and higher (please see Service Page 1) safety func-
tion is installed, which controls swivelled in vacuum chamber. With press-
ing the button for 3 seconds you can disable warning W077, if you want
to open the sealing rod for cleaning empty hot crucible.
Vacuum
Pump
Switching on and off the vacuum pump, when the pump is connected to
the backside socket from the machine.
Automatic
Start
With this button the selected casting program starts and its parameter will
be executed from the memory. Starts pausing program (LED is blinking).
Automatic
Stop
Pausing an automatic program if pressed once and stops the program if
you press it twice.