11
Explorer
Pro
3.3.5
Calibration Test
Calibration test feature allows a check of a known calibration mass against the last stored calibration information in the balance.
The sample displays shown are for a 12 kilogram balance.
Go Back
Help
Enter
On/Off
PROCEDURE
• Press
arrow button to select MAIN MENU.
• Press
Enter
button.
• Using the arrow buttons, select CALIBRATION.
• Press
Enter
button.
• Press
or
arrow button and select CALIBRATION
TEST.
• Press
Enter
button.
• Clear the pan and press
Enter
button.
• Place the indicated mass value on the pan and press
Enter
button.
• Display indicates if calibration was successful and the
difference between the last calibration.
• Remove masses from the pan.
• Press any button, balance returns to weighing mode.
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CALIBRATION TEST
PLEASE CLEAR THE PAN
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CALIBRATION TEST
PLEASE WAIT . . .
Go Back
Help
Enter
On/Off
0.0
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CALIBRATION TEST
PLEASE PUT: 12000.00 G ON PAN.
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
CALIBRATION TEST
PLEASE WAIT . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
ENTER USER CAL VALUE
0
CALIBRATION TEST
SPAN CAL SUCCESSFULLY DONE!
THE LAST DIFFERENCE BETWEEN THIS CAL
AND LAST CAL IS:
0.1 GRAMS
PRESS ANY KEY TO EXIT . . .
0.0
WEIGHT
0%
100%
CONTRAST
GRAMS
STABLE
MOVE HIGHLIGHT
CONTRAST
MAIN MENU
Summary of Contents for BH350
Page 64: ...56 Explorer Pro NOTES ...