9
Explorer
Pro
Go Back
Help
Enter
On/Off
Go Back
Help
Enter
On/Off
PROCEDURE
• Press
arrow button to select MAIN MENU.
• Press
Enter
button.
• Using the arrow buttons, select CALIBRATION.
• Press
Enter
button.
• Press
or
arrow button and select LINEARITY.
• Press
Enter
button, LINEARITY CALIBRATION is displayed.
• Clear pan and press
Enter
button.
• Place indicated mass value on pan. This will be one half
of the total capacity of the balance, then press
Enter
button.
• Place indicated mass value on pan. This will be the total
capacity of the balance, then press
Enter
button.
PLEASE WAIT is displayed followed by the actual weight value
placed on the pan which is the maximum capacity of the
balance.
• Remove the masses from the pan. The balance is now
calibrated.
NOTE:
The samples shown on the displays were for an 12 kg
balance.
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
LINEARITY CALIBRATION
PLEASE CLEAR THE PAN
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
LINEARITY CALIBRATION
PLEASE WAIT . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
LINEARITY CALIBRATION
PLEASE WAIT . . .
3.3.3
Linearity Calibration
Linearity calibration utilizes three calibration points, one at zero, center span and full span. This method minimizes deviation
between actual and displayed weights within the balance's weighing range. Three weight values are used; zero, a weight value
at midpoint of the balance's weighing range and a weight value at or near the specified capacity.
MAIN MENU
CHANGE UNITS
SET BALANCE
LIBRARY
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
LINEARITY CALIBRATION
PLEASE PUT 6000.00 G ON PAN
AND PRESS < ENTER > . . .
MAIN MENU
CHANGE UNITS
SET BALANCE
RETURN TO
WEIGH
CUSTOM MENU
CHANGE MODE
CALIBRATION
CALIBRATE
SPAN
LINPLEASEERARITY
USER
CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH
LINEARITY CALIBRATION
PLEASE PUT 12000.00 G ON PAN
AND PRESS < ENTER > . . .
0.0
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
MAIN MENU
CONTRAST
12000.0
WEIGHT
0%
100%
GRAMS
STABLE
MOVE HIGHLIGHT
CONTRAST
MAIN MENU